Guidance and manufacturer’s declaration – electromagnetic 

immunity

The ProCair Trio control unit is intended for use in the electromagnetic environment speci

fi

 ed below. The 

customer or the user of the ProCair Trio control unit should ensure that it is used in such an environment.

EMI/EMC Statement and Manufacturer’s 

Declaration

Emission test

Compliance

Electromagnetic environment-guidance

RF emissions CISPR 11

Group 1

The ProCair Trio control unit uses RF energy only for its internal 
function. Therefore, its RF emissions are very low and are not likely 
to cause any interference in nearby electronic equipment.

RF emissions CISPR 11

Class B

The ProCair Trio control unit is suitable for use in all 
establishments, including domestic establishments and those 
directly connected to the public low-voltage power supply network 
that supplies buildings used for domestic purposes.

Harmonic emissions 
IEC 61000-3-2

Class A (230V)

Voltage 

fl

 uctuations /

fl

 icker

emissions IEC 61000-3-3

Compliance (230V)

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment-guidance

Electrostatic

discharge(ESD) IEC

61000-4-2

Contact ±8 kV

Contact ±8 kV

Floors should be wood, concrete or ceramic 
tile. If  

fl

  oors are covered with synthetic 

material, the relative humidity should be at 
least 30%

Air ±2, ±4, ±8, ±15 kV

Air ±2, ±4, ±8, ±15 kV

Electrical fast 

transient/burst 

IEC 61000-4-4

± 2kV for power

supply lines

± 2kV for power 

supply lines

Mains power quality should be that of a typical 
home healthcare and professional healthcare 
environment.

± 1kV for input/

output lines

Not applicable

Surge

IEC 61000-4-5

± 0.5, ±1 kV line(s) to

line(s)

± 0.5, ±1 kV line(s) to

line(s)

Mains power quality should be that of a typical 
home healthcare and professional healthcare 
environment.

± 0.5, ±1, ± 2kV line(s) 

to earth

Not applicable

Voltage Dips, short 

interruptions and 

voltage variations on 

power supply input 

lines

IEC 61000-4-11

Voltage dips

0 % UT; 0.5 cycle
0 % UT; 0.1 cycle

70 % UT; 25/30 cycles

Voltage dips

0 % UT; 0.5 cycle

0 % UT; 0.1 cycle

70 % UT; 25/30 cycles

Mains power quality should be that of a 
typical home healthcare and professional 
healthcare environment. If the user of the 
ProCair Trio control unit requires continued 
operation during power main interruptions, it 
is recommended that the ProCair Trio control 
unit be powered from an uninterruptible power 
supply or a battery.

Voltage interruptions 
0 % UT; 250/300 cycle

Voltage interruptions 
0 % UT; 250/300 cycle

Power frequency

(50/60 Hz) magnetic

fi

  eld IEC 61000-4-8

30 A/m

30 A/m

Power frequency magnetic 

fi

  elds should be 

at levels characteristic of a typical location in 
a typical home healthcare and professional 
healthcare environment.

50 Hz or 60 Hz

50 Hz

Summary of Contents for APMPC-R05

Page 1: ...pressure care and patient handling specialists novis com au 1300 738 885 Static Mode Programming Option ProCairTrio UserGuide...

Page 2: ...dreservestherightto amendspecificationspresentedinthisguide Informationcorrectattimeofproduction September2020 2020NovisHealthcare Allrightsreserved Definition of Symbols Used Thefollowingsymbolsmayap...

Page 3: ...tion 9 Patient Set Up 12 Operation Control Panel 13 Operation Mode 16 Operation Pull Cap CPR Release 17 Operation Deflation Storage 18 Care and Cleaning 19 Troubleshooting 21 Technical Specifications...

Page 4: ...atient ssurface tissue maintainingandstimulatingtheflowof bloodandlymphaticfluidsthroughthesetissues toprovideessentialoxygenandremovemetabolic waste Thesystemconsistsofthefollowingcomponents Mattress...

Page 5: ...areHomehealthcareand Professionalhealthcare Working Environment ControlUnit Temperature 15 Cto35 C 59 Fto95 F Humidity 30 to75 non condensing Shipping Storage Environment ControlUnit Temperature 5 Cto...

Page 6: ...ndividuals Thisdevicecanbeusedinhomehealthcareand professionalhealthcareenvironment Medicalelectricalequipmentneedsspecial precautionsregardingEMCandneedstobe installedaccordingtotheEMCinformation pro...

Page 7: ...ely Avoidblockingtheairintakesofthecontrol unit locatedattherearoftheunit Donotplace itemssuchasblanketsoverthecontrolunit Bedframesusedwiththesystemscanvary greatlydependingonthespecifichealthcare se...

Page 8: ...fsmoking materialsoropenflame airflowingthrough themattresswillsupportcombustion Donotopenthecontrolunit riskof electricalshock Referservicingtoqualified servicepersonnel Disposal Disposeofallcomponen...

Page 9: ...wardsthebaseofthebed 2 Attachto the bed bysecuringthe adjustable straps locatedontheundersideofthe mattressbaseundereachbedend Ona profilingbed securethestrapsaroundthe moveablesectionsofthebase Ensur...

Page 10: ...dhanging hooks Pullthehooksbytherubbertabsto preventaccidentallytrappingyourfingers Ensureitissecurebeforeuse failuretodoso couldresultinequipmentdamage 6 Connecttheumbilicalconnectorstothe socketsons...

Page 11: ...flash green Allow upto 20minutes for completeinflation Donotlieaperson orany weight onthemattressduringinitialinflation Donotplaceaperson oranyitem onthe mattressduringinitialinflation 9 Wheninitialin...

Page 12: ...to feelthe cellsforinflation 2 Withthepatientlyingsupine unzipone sideofthe top cover justpastsacralregion lowerspine 3 Slideyourhandunderneaththepatientand feelfor a deflatedcellunderthepatient s lo...

Page 13: ...therapyanduseradjustablepressuresettings TheselectedmodeisindicatedbyanLEDindicator D AlarmIndicator ThethreeamberLEDslightupin5different patternstoindicatedifferentfaultsthesystem maybeexperiencing...

Page 14: ...Thisfeaturehelpstopreventtheriskofbottoming out Agreen AUTOFOWLER lightilluminateswhenthis functionisactive Itisautomaticallydeactivated whenthemattressbackrestisprofiledbelow5 HPower StandbyandServic...

Page 15: ...ruptionto therapy due to power failure Note Itwilltake16 20hourstofullychargethe battery K BedExitAlarm PressingthebuttonturnsBedExitAlarmonandoff TheONlightabovebuttonilluminatestoindicatethe functio...

Page 16: ...icallydisengage Backup Battery Thelithium ionbackupbatteryisdesignedto providetemporaryuninterruptedtherapywhen mainspowerisunavailableegduringpatient transferorinapoweroutage Thecontrolunitautomatica...

Page 17: ...ksareheard Airwillremaininthe systemforupto24hours dependingonpatient andenvironmentalcircumstances Specialised Heel Therapy ThespecialisedHeelCairzoneenablestheseven narrowalternatingaircellsinthehee...

Page 18: ...sedfromthesystem detachthemattressfromthebedby unfasteningthestraps thenfoldandroll themattressfromheadendtofootendfor storage 6 Returnallitemstothecustomcarrybagfor safekeeping CPRRELEASE POWER STAND...

Page 19: ...hadryclothtoremoveexcess moisture Asoftbristlednylonbrushcanbeusedto gentlycleancrevices Ensurethecontrolunitisdisconnectedfrom mainspowerbeforecleaning Donotspraydisinfectantdirectlyontothe controlun...

Page 20: ...lunitfacedownonasoft flat surfacewithbackpaneluppermost useasoft clothtopreventscratches 3 Useasmallscrewdrivertocarefullyremovethe airfiltercover Cleanthedustfromthefilteror discardandreplacewithanew...

Page 21: ...for kinks obstructions or damage Undo any kinks and obstructions Air cells have failed to reach operating pressure after turning on Turn of control unit and disconnect from power Reconnect to power a...

Page 22: ...ate no display lights Check control unit is connected to the mains power supply Check for loose power cord connection and ensure main power is switched on Check the fuses in control unit Replace if ne...

Page 23: ...ENT 0 4 0 2A FUSE RATING T2AL250V MIN MAX PRESSURE 20 60mmHg 6mmHg FLOW RATE Max10L min PROTECTION TYPE ClassIITypeBF INGRESS PROTECTION RATING IP21 CONTROL UNIT DIMENSION HEIGHT 326mm WIDTH 275mm DEP...

Page 24: ...dsThefirst digit indicates thelevelofprotectionthattheenclosure providesagainst accessto hazardous parts e g electricalconductors moving parts andtheingressofsolidforeignobjects SecondDigit LiquidsPro...

Page 25: ...nsequipment Fortransmittersratedatamaximumoutputpowernotlistedabove therecommendedseparationdistanced inmetres m canbeestimatedusingtheequationapplicabletothefrequencyofthetransmitter wherePisthe maxi...

Page 26: ...ative humidity should be at least 30 Air 2 4 8 15 kV Air 2 4 8 15 kV Electrical fast transient burst IEC 61000 4 4 2kV for power supply lines 2kV for power supply lines Mains power quality should be t...

Page 27: ...eryInUse LEDisnormal duetopowerissuppliedbyinternalbatteryduringpoweroutage Thebackupbatterycouldbeusedforupto3hours EMI EMC Statement and Manufacturer s Declaration Immunitytest IEC60601testlevel Com...

Page 28: ...mendedseparationdistance calculatedfromtheequationapplicable tothefrequencyofthetransmitter Radiated RF IEC 61000 4 3 10V m 80MHz 2 7GHz 80 AMat1kHz 10V m 80MHz 2 7GHz 80 AMat1kHz Recommendedseparatio...

Page 29: ...on Pleasedonotuseanyothercablesoraccessoriesnotapprovedbythemanufacturerinthismanualtoavoidnegative influenceonelectromagneticcompatibility EMI EMC Statement and Manufacturer s Declaration Testfrequen...

Page 30: ...ished good of the same type rather than being repaired Refurbished parts may be used to repair goods NovisHealthcarewillnotbeheldresponsibleforanyrepair otherthanthosecarriedoutbyitoroneofitsAuthorise...

Page 31: ...31 ProCair Trio User Guide Notes...

Page 32: ...Pressure care and patient handling specialists 06102020F 1300738885 novis novis com au 12 12MarsRd LaneCoveWest NSW2066...

Reviews: