Novis APMPC-R05 User Manual Download Page 23

23

ProCair Trio User Guide

Technical Speci

fi

 cations

SYSTEM

MODEL

ProCair Trio

ProCair Trio King Single

SYSTEM CODE

APMPC-R05

APMPC-R05K

CAPACITY

40 - 250 kg

NO OF CELLS

20, including 3 static head cells and 7 disconnectable heel cells.

All alternating cells include a static lower chamber (cell-in-cell)

COMPLIANCE

IEC60601-1, IEC60601-1-2 and IEC60601-1-11

ARTG

289458

CONTROL UNIT

PART NO.

APMPC-CU05

CONTROL SYSTEM

Digital micro controller

CYCLE TIME

15 minutes (

fi

 xed)

SUPPLY VOLTAGE

AC100 - 240V / 50Hz-60Hz

MAXIMUM CURRENT

0.4 ~0.2 A

FUSE RATING

T2AL 250V

MIN / MAX PRESSURE

20 ~ 60 mmHg +/- 6 mmHg

FLOW RATE

Max 10 L/min

PROTECTION TYPE

Class II Type BF

INGRESS PROTECTION RATING

IP21

CONTROL UNIT 
DIMENSION

HEIGHT

326 mm

WIDTH

275 mm

DEPTH

133 mm

WEIGHT

5 kg

MATTRESS
DIMENSIONS

LENGTH 

2000 mm

2000 mm

WIDTH

880 mm

1050 mm

HEIGHT

200 mm

200 mm

WEIGHT

14.5 kg

16 kg

MATERIAL

TOP COVER

PU laminated nylon

BASE COVER

TPU-laminated PU, fully sealed

AIR CELL

TPU

OPERATING 
ENVIRONMENT

AIR HUMIDITY

OPERATION

30% to 75% non-condensing

STORAGE

30% to 90% non-condensing

MATRESS STORAGE

10% ~90%

AMBIENT 

TEMPERATURE

OPERATION

15° C to 35° C

STORAGE

5° C to 60° C

MATRESS STORAGE

-10°C ~ 40°C

ATMOSPHERIC PRESSURE RANGE

700 hPa to 1060 hPa

ALTITUDE

-310 metres to 3000 metres

Summary of Contents for APMPC-R05

Page 1: ...pressure care and patient handling specialists novis com au 1300 738 885 Static Mode Programming Option ProCairTrio UserGuide...

Page 2: ...dreservestherightto amendspecificationspresentedinthisguide Informationcorrectattimeofproduction September2020 2020NovisHealthcare Allrightsreserved Definition of Symbols Used Thefollowingsymbolsmayap...

Page 3: ...tion 9 Patient Set Up 12 Operation Control Panel 13 Operation Mode 16 Operation Pull Cap CPR Release 17 Operation Deflation Storage 18 Care and Cleaning 19 Troubleshooting 21 Technical Specifications...

Page 4: ...atient ssurface tissue maintainingandstimulatingtheflowof bloodandlymphaticfluidsthroughthesetissues toprovideessentialoxygenandremovemetabolic waste Thesystemconsistsofthefollowingcomponents Mattress...

Page 5: ...areHomehealthcareand Professionalhealthcare Working Environment ControlUnit Temperature 15 Cto35 C 59 Fto95 F Humidity 30 to75 non condensing Shipping Storage Environment ControlUnit Temperature 5 Cto...

Page 6: ...ndividuals Thisdevicecanbeusedinhomehealthcareand professionalhealthcareenvironment Medicalelectricalequipmentneedsspecial precautionsregardingEMCandneedstobe installedaccordingtotheEMCinformation pro...

Page 7: ...ely Avoidblockingtheairintakesofthecontrol unit locatedattherearoftheunit Donotplace itemssuchasblanketsoverthecontrolunit Bedframesusedwiththesystemscanvary greatlydependingonthespecifichealthcare se...

Page 8: ...fsmoking materialsoropenflame airflowingthrough themattresswillsupportcombustion Donotopenthecontrolunit riskof electricalshock Referservicingtoqualified servicepersonnel Disposal Disposeofallcomponen...

Page 9: ...wardsthebaseofthebed 2 Attachto the bed bysecuringthe adjustable straps locatedontheundersideofthe mattressbaseundereachbedend Ona profilingbed securethestrapsaroundthe moveablesectionsofthebase Ensur...

Page 10: ...dhanging hooks Pullthehooksbytherubbertabsto preventaccidentallytrappingyourfingers Ensureitissecurebeforeuse failuretodoso couldresultinequipmentdamage 6 Connecttheumbilicalconnectorstothe socketsons...

Page 11: ...flash green Allow upto 20minutes for completeinflation Donotlieaperson orany weight onthemattressduringinitialinflation Donotplaceaperson oranyitem onthe mattressduringinitialinflation 9 Wheninitialin...

Page 12: ...to feelthe cellsforinflation 2 Withthepatientlyingsupine unzipone sideofthe top cover justpastsacralregion lowerspine 3 Slideyourhandunderneaththepatientand feelfor a deflatedcellunderthepatient s lo...

Page 13: ...therapyanduseradjustablepressuresettings TheselectedmodeisindicatedbyanLEDindicator D AlarmIndicator ThethreeamberLEDslightupin5different patternstoindicatedifferentfaultsthesystem maybeexperiencing...

Page 14: ...Thisfeaturehelpstopreventtheriskofbottoming out Agreen AUTOFOWLER lightilluminateswhenthis functionisactive Itisautomaticallydeactivated whenthemattressbackrestisprofiledbelow5 HPower StandbyandServic...

Page 15: ...ruptionto therapy due to power failure Note Itwilltake16 20hourstofullychargethe battery K BedExitAlarm PressingthebuttonturnsBedExitAlarmonandoff TheONlightabovebuttonilluminatestoindicatethe functio...

Page 16: ...icallydisengage Backup Battery Thelithium ionbackupbatteryisdesignedto providetemporaryuninterruptedtherapywhen mainspowerisunavailableegduringpatient transferorinapoweroutage Thecontrolunitautomatica...

Page 17: ...ksareheard Airwillremaininthe systemforupto24hours dependingonpatient andenvironmentalcircumstances Specialised Heel Therapy ThespecialisedHeelCairzoneenablestheseven narrowalternatingaircellsinthehee...

Page 18: ...sedfromthesystem detachthemattressfromthebedby unfasteningthestraps thenfoldandroll themattressfromheadendtofootendfor storage 6 Returnallitemstothecustomcarrybagfor safekeeping CPRRELEASE POWER STAND...

Page 19: ...hadryclothtoremoveexcess moisture Asoftbristlednylonbrushcanbeusedto gentlycleancrevices Ensurethecontrolunitisdisconnectedfrom mainspowerbeforecleaning Donotspraydisinfectantdirectlyontothe controlun...

Page 20: ...lunitfacedownonasoft flat surfacewithbackpaneluppermost useasoft clothtopreventscratches 3 Useasmallscrewdrivertocarefullyremovethe airfiltercover Cleanthedustfromthefilteror discardandreplacewithanew...

Page 21: ...for kinks obstructions or damage Undo any kinks and obstructions Air cells have failed to reach operating pressure after turning on Turn of control unit and disconnect from power Reconnect to power a...

Page 22: ...ate no display lights Check control unit is connected to the mains power supply Check for loose power cord connection and ensure main power is switched on Check the fuses in control unit Replace if ne...

Page 23: ...ENT 0 4 0 2A FUSE RATING T2AL250V MIN MAX PRESSURE 20 60mmHg 6mmHg FLOW RATE Max10L min PROTECTION TYPE ClassIITypeBF INGRESS PROTECTION RATING IP21 CONTROL UNIT DIMENSION HEIGHT 326mm WIDTH 275mm DEP...

Page 24: ...dsThefirst digit indicates thelevelofprotectionthattheenclosure providesagainst accessto hazardous parts e g electricalconductors moving parts andtheingressofsolidforeignobjects SecondDigit LiquidsPro...

Page 25: ...nsequipment Fortransmittersratedatamaximumoutputpowernotlistedabove therecommendedseparationdistanced inmetres m canbeestimatedusingtheequationapplicabletothefrequencyofthetransmitter wherePisthe maxi...

Page 26: ...ative humidity should be at least 30 Air 2 4 8 15 kV Air 2 4 8 15 kV Electrical fast transient burst IEC 61000 4 4 2kV for power supply lines 2kV for power supply lines Mains power quality should be t...

Page 27: ...eryInUse LEDisnormal duetopowerissuppliedbyinternalbatteryduringpoweroutage Thebackupbatterycouldbeusedforupto3hours EMI EMC Statement and Manufacturer s Declaration Immunitytest IEC60601testlevel Com...

Page 28: ...mendedseparationdistance calculatedfromtheequationapplicable tothefrequencyofthetransmitter Radiated RF IEC 61000 4 3 10V m 80MHz 2 7GHz 80 AMat1kHz 10V m 80MHz 2 7GHz 80 AMat1kHz Recommendedseparatio...

Page 29: ...on Pleasedonotuseanyothercablesoraccessoriesnotapprovedbythemanufacturerinthismanualtoavoidnegative influenceonelectromagneticcompatibility EMI EMC Statement and Manufacturer s Declaration Testfrequen...

Page 30: ...ished good of the same type rather than being repaired Refurbished parts may be used to repair goods NovisHealthcarewillnotbeheldresponsibleforanyrepair otherthanthosecarriedoutbyitoroneofitsAuthorise...

Page 31: ...31 ProCair Trio User Guide Notes...

Page 32: ...Pressure care and patient handling specialists 06102020F 1300738885 novis novis com au 12 12MarsRd LaneCoveWest NSW2066...

Reviews: