ProCair Trio User Guide

Cleaning and Infection Control

 It is recommended that the ProCair Trio 
system is cleaned every two weeks if in 
constant use.

Top Cover Cleaning

Unzip and remove the top cover from the base 
before washing (refer page 18 for instructions).

For basic care and cleaning, wipe down with warm 
water containing PH neutral detergent. The top 
cover can also be machine washed at a maximum 
of 95° C (203° F) using neutral detergents.

Refer to the top cover wash tag for detailed 
cleaning instructions.

Do not use system without top cover.

Base and Cells Cleaning

Swab the mattress base and cells with a solution 
of sodium hypochlorite or similar (up to 10,000 
ppm available chlorine). Dry thoroughly before 
reassembly.

Do not machine wash or tumble dry the air cells or 
mattress base.

If cleaning or disinfection is required, do not 
allow 

fl

  uid to enter air cells and air hoses.

Control Unit Cleaning

Disconnect control unit from mains power before 
cleaning. Gently wipe down the external case with 
a soft cloth.

Soak the cloth in warm water containing mild pH 
neutral detergent, and wring any excess water 
before gently wiping all external controls. Repeat 
the process with a dry cloth to remove excess 
moisture. A soft bristled nylon brush can be used to 
gently clean crevices.

Ensure the control unit is disconnected from 
mains power before cleaning.

Do not spray disinfectant directly on to the 
control unit, or immerse the unit in water or 
other 

fl

 uid.

Disinfection

The mattress, top cover and control unit may 
be decontaminated by a solution of sodium 
hypochlorite or similar (up to 10,000 ppm available 
chlorine). Dry thoroughly before use.

For infection control, swab with a solution of 
sodium hypochlorite or similar (up to 10,000 
ppm available chlorine). Dry thoroughly before 
reattaching and use.

Care and Cleaning

To prevent cross contamination, the mattress should be examined and disinfected between patient use. 

Clean the mattress in accordance with local infection control policy and government regulations. Failure 
to do so could cause patient or personal injury.

The mattress is not protected against excessive amounts of 

fl

  uid. Do not immerse the control unit in 

fl

 uid.

Switch off and disconnect the control unit from mains power supply before cleaning. Failure to do so 
could result in equipment damage or electric shock.

Do not use high temperature autoclave steam cleaning devices or phenolic based products for cleaning. 
This could result in damage to the equipment and may result in damage to the polyurethane coating, or 
negate the biocompatibility properties of the fabric.

Summary of Contents for APMPC-R05

Page 1: ...pressure care and patient handling specialists novis com au 1300 738 885 Static Mode Programming Option ProCairTrio UserGuide...

Page 2: ...dreservestherightto amendspecificationspresentedinthisguide Informationcorrectattimeofproduction September2020 2020NovisHealthcare Allrightsreserved Definition of Symbols Used Thefollowingsymbolsmayap...

Page 3: ...tion 9 Patient Set Up 12 Operation Control Panel 13 Operation Mode 16 Operation Pull Cap CPR Release 17 Operation Deflation Storage 18 Care and Cleaning 19 Troubleshooting 21 Technical Specifications...

Page 4: ...atient ssurface tissue maintainingandstimulatingtheflowof bloodandlymphaticfluidsthroughthesetissues toprovideessentialoxygenandremovemetabolic waste Thesystemconsistsofthefollowingcomponents Mattress...

Page 5: ...areHomehealthcareand Professionalhealthcare Working Environment ControlUnit Temperature 15 Cto35 C 59 Fto95 F Humidity 30 to75 non condensing Shipping Storage Environment ControlUnit Temperature 5 Cto...

Page 6: ...ndividuals Thisdevicecanbeusedinhomehealthcareand professionalhealthcareenvironment Medicalelectricalequipmentneedsspecial precautionsregardingEMCandneedstobe installedaccordingtotheEMCinformation pro...

Page 7: ...ely Avoidblockingtheairintakesofthecontrol unit locatedattherearoftheunit Donotplace itemssuchasblanketsoverthecontrolunit Bedframesusedwiththesystemscanvary greatlydependingonthespecifichealthcare se...

Page 8: ...fsmoking materialsoropenflame airflowingthrough themattresswillsupportcombustion Donotopenthecontrolunit riskof electricalshock Referservicingtoqualified servicepersonnel Disposal Disposeofallcomponen...

Page 9: ...wardsthebaseofthebed 2 Attachto the bed bysecuringthe adjustable straps locatedontheundersideofthe mattressbaseundereachbedend Ona profilingbed securethestrapsaroundthe moveablesectionsofthebase Ensur...

Page 10: ...dhanging hooks Pullthehooksbytherubbertabsto preventaccidentallytrappingyourfingers Ensureitissecurebeforeuse failuretodoso couldresultinequipmentdamage 6 Connecttheumbilicalconnectorstothe socketsons...

Page 11: ...flash green Allow upto 20minutes for completeinflation Donotlieaperson orany weight onthemattressduringinitialinflation Donotplaceaperson oranyitem onthe mattressduringinitialinflation 9 Wheninitialin...

Page 12: ...to feelthe cellsforinflation 2 Withthepatientlyingsupine unzipone sideofthe top cover justpastsacralregion lowerspine 3 Slideyourhandunderneaththepatientand feelfor a deflatedcellunderthepatient s lo...

Page 13: ...therapyanduseradjustablepressuresettings TheselectedmodeisindicatedbyanLEDindicator D AlarmIndicator ThethreeamberLEDslightupin5different patternstoindicatedifferentfaultsthesystem maybeexperiencing...

Page 14: ...Thisfeaturehelpstopreventtheriskofbottoming out Agreen AUTOFOWLER lightilluminateswhenthis functionisactive Itisautomaticallydeactivated whenthemattressbackrestisprofiledbelow5 HPower StandbyandServic...

Page 15: ...ruptionto therapy due to power failure Note Itwilltake16 20hourstofullychargethe battery K BedExitAlarm PressingthebuttonturnsBedExitAlarmonandoff TheONlightabovebuttonilluminatestoindicatethe functio...

Page 16: ...icallydisengage Backup Battery Thelithium ionbackupbatteryisdesignedto providetemporaryuninterruptedtherapywhen mainspowerisunavailableegduringpatient transferorinapoweroutage Thecontrolunitautomatica...

Page 17: ...ksareheard Airwillremaininthe systemforupto24hours dependingonpatient andenvironmentalcircumstances Specialised Heel Therapy ThespecialisedHeelCairzoneenablestheseven narrowalternatingaircellsinthehee...

Page 18: ...sedfromthesystem detachthemattressfromthebedby unfasteningthestraps thenfoldandroll themattressfromheadendtofootendfor storage 6 Returnallitemstothecustomcarrybagfor safekeeping CPRRELEASE POWER STAND...

Page 19: ...hadryclothtoremoveexcess moisture Asoftbristlednylonbrushcanbeusedto gentlycleancrevices Ensurethecontrolunitisdisconnectedfrom mainspowerbeforecleaning Donotspraydisinfectantdirectlyontothe controlun...

Page 20: ...lunitfacedownonasoft flat surfacewithbackpaneluppermost useasoft clothtopreventscratches 3 Useasmallscrewdrivertocarefullyremovethe airfiltercover Cleanthedustfromthefilteror discardandreplacewithanew...

Page 21: ...for kinks obstructions or damage Undo any kinks and obstructions Air cells have failed to reach operating pressure after turning on Turn of control unit and disconnect from power Reconnect to power a...

Page 22: ...ate no display lights Check control unit is connected to the mains power supply Check for loose power cord connection and ensure main power is switched on Check the fuses in control unit Replace if ne...

Page 23: ...ENT 0 4 0 2A FUSE RATING T2AL250V MIN MAX PRESSURE 20 60mmHg 6mmHg FLOW RATE Max10L min PROTECTION TYPE ClassIITypeBF INGRESS PROTECTION RATING IP21 CONTROL UNIT DIMENSION HEIGHT 326mm WIDTH 275mm DEP...

Page 24: ...dsThefirst digit indicates thelevelofprotectionthattheenclosure providesagainst accessto hazardous parts e g electricalconductors moving parts andtheingressofsolidforeignobjects SecondDigit LiquidsPro...

Page 25: ...nsequipment Fortransmittersratedatamaximumoutputpowernotlistedabove therecommendedseparationdistanced inmetres m canbeestimatedusingtheequationapplicabletothefrequencyofthetransmitter wherePisthe maxi...

Page 26: ...ative humidity should be at least 30 Air 2 4 8 15 kV Air 2 4 8 15 kV Electrical fast transient burst IEC 61000 4 4 2kV for power supply lines 2kV for power supply lines Mains power quality should be t...

Page 27: ...eryInUse LEDisnormal duetopowerissuppliedbyinternalbatteryduringpoweroutage Thebackupbatterycouldbeusedforupto3hours EMI EMC Statement and Manufacturer s Declaration Immunitytest IEC60601testlevel Com...

Page 28: ...mendedseparationdistance calculatedfromtheequationapplicable tothefrequencyofthetransmitter Radiated RF IEC 61000 4 3 10V m 80MHz 2 7GHz 80 AMat1kHz 10V m 80MHz 2 7GHz 80 AMat1kHz Recommendedseparatio...

Page 29: ...on Pleasedonotuseanyothercablesoraccessoriesnotapprovedbythemanufacturerinthismanualtoavoidnegative influenceonelectromagneticcompatibility EMI EMC Statement and Manufacturer s Declaration Testfrequen...

Page 30: ...ished good of the same type rather than being repaired Refurbished parts may be used to repair goods NovisHealthcarewillnotbeheldresponsibleforanyrepair otherthanthosecarriedoutbyitoroneofitsAuthorise...

Page 31: ...31 ProCair Trio User Guide Notes...

Page 32: ...Pressure care and patient handling specialists 06102020F 1300738885 novis novis com au 12 12MarsRd LaneCoveWest NSW2066...

Reviews: