13
ProCair Trio User Guide
Operation - Control Panel
A
Pressure Setting - Fine Tuning
The control unit constantly monitors mattress
loading to automatically adjust to the optimal
pressure setting.
For individual comfort or therapy needs, slight
adjustments to the automatic pressure setting can
be made. These arrows allow optimal pressure
setting to be
fi
ne tuned softer or
fi
rmer.
“Normal” light illuminates to indicate standard
weight detection is functioning.
The control unit would not allow the pressure
setting to go below the safe lower limit for the
patient according to mattress load
B
Pressure Setting Display
Displays pressure setting range selected.
C
Mode
Pressing the Mode button toggles between
Alternate mode and Inde
fi
nite Static mode (all cells
in
fl
ated with no dynamic alternation) Inde
fi
nite
Static mode allows for constant low pressure
therapy and user adjustable pressure settings.
The selected mode is indicated by an LED indicator.
D
Alarm Indicator
The three amber LEDs light up in 5 different
patterns to indicate different faults the system
may be experiencing, this is accompanied by an
audible beeping signal. Refer to the alarm guide
on the left side on the control panel to look up the
fault indicated, also refer to the corresponding
troubleshooting guide (page 21) for detail, alarms
include:
LOW PRESSURE - possible loss of pressure
CPR RELEASE
POWER
STANDBY
ALTERNATE
STATIC
PRESSURE
SETTING
ProCair Trio
LOCATED AT
MATTRESS
HEAD END -
TWIST TO OPEN
NORMAL
<30%
FLASHING: CHARGING
FLASHING: LOW BATT
FLASHING
SERVICE DUE
LOW PRESSURE
SOFTER
NORMAL
FIRMER
HIGH PRESSURE
STARTUP FAILURE
ALTERNATION
FAILURE
SENSOR
DISCONNECTED
ALARM GUIDE
LOCKED
ON
AUTO
FOWLER
BED EXIT ALARM
see alarm guide on left
ALARM INDICATORS
BATT.
IN USE
A
B
C
D
D
E
F
G
J
H
I
K
Press the left arrow
to select a softer
pressure pressure
range
Press the right
arrow to select a
fi
rmer pressure
pressure range