Novis APMPC-R05 User Manual Download Page 12

12

Patient Set Up

Once initial in

fl

 ation is

complete, a patient may be
placed on the system.

1

Once the mattress is fully in

fl

 ated, bedding 

can be replaced. 
Secure sheets loosely enough to ensure 
they do not interfere with cell alternation.

2

Place the patient on to the ProCair Trio 

mattress. 
The system will automatically set an 
optimum pressure for the patient’s weight 
(weight range from 40 to 250 kg) and will 
continuously alternate over a 15 minute 
cycle.

3

Perform a ‘bottoming out’ test (a test to 

ensure the patient is adequately suspended 
away from the base).

“Bottoming Out” Test

1    

Check system is in alternation mode by 
ensuring the ALTERNATE indicator is 

illuminated ,and that three out of every four 
air cells is in

fl

  ated while the fourth cell is 

de

fl

 ated. 

You may need to unzip the cover to feel the 
cells for in

fl

 ation.

2

With the patient lying supine, unzip one 

side of the top cover just past sacral region 
(lower spine).

3

Slide your hand underneath the patient and 

feel for a de

fl

  ated cell under the patient’s 

lower spine. The inner static cell will remain 
in

fl

  ated, however your hand should easily 

slide between patient and base. 

4

If your hand can pass under the patient, 

then patient is adequately suspended. If 
not, manually adjust pressure to ‘

fi

 rmer’ 

and wait at least one cycle (15 minutes) 
for pressure to increase before repeating 
step 3. 
If manual pressure adjustment fails, 
press Max In

fl

  ate to force mattress to 

full in

fl

  ation. Wait at least one cycle (15 

minutes) for pressure to reach maximum 
pressure, then press Alt to return to an 
alternation cycle. Wait at least one more 
cycle (15 minutes) for pressure to increase 
before repeating step 3.

5

We recommend repeating the Bottoming 

Out test at least 15 minutes after any 
manual pressure readjustment and 
regularly every 2 hours during the system in 
operation.

Summary of Contents for APMPC-R05

Page 1: ...pressure care and patient handling specialists novis com au 1300 738 885 Static Mode Programming Option ProCairTrio UserGuide...

Page 2: ...dreservestherightto amendspecificationspresentedinthisguide Informationcorrectattimeofproduction September2020 2020NovisHealthcare Allrightsreserved Definition of Symbols Used Thefollowingsymbolsmayap...

Page 3: ...tion 9 Patient Set Up 12 Operation Control Panel 13 Operation Mode 16 Operation Pull Cap CPR Release 17 Operation Deflation Storage 18 Care and Cleaning 19 Troubleshooting 21 Technical Specifications...

Page 4: ...atient ssurface tissue maintainingandstimulatingtheflowof bloodandlymphaticfluidsthroughthesetissues toprovideessentialoxygenandremovemetabolic waste Thesystemconsistsofthefollowingcomponents Mattress...

Page 5: ...areHomehealthcareand Professionalhealthcare Working Environment ControlUnit Temperature 15 Cto35 C 59 Fto95 F Humidity 30 to75 non condensing Shipping Storage Environment ControlUnit Temperature 5 Cto...

Page 6: ...ndividuals Thisdevicecanbeusedinhomehealthcareand professionalhealthcareenvironment Medicalelectricalequipmentneedsspecial precautionsregardingEMCandneedstobe installedaccordingtotheEMCinformation pro...

Page 7: ...ely Avoidblockingtheairintakesofthecontrol unit locatedattherearoftheunit Donotplace itemssuchasblanketsoverthecontrolunit Bedframesusedwiththesystemscanvary greatlydependingonthespecifichealthcare se...

Page 8: ...fsmoking materialsoropenflame airflowingthrough themattresswillsupportcombustion Donotopenthecontrolunit riskof electricalshock Referservicingtoqualified servicepersonnel Disposal Disposeofallcomponen...

Page 9: ...wardsthebaseofthebed 2 Attachto the bed bysecuringthe adjustable straps locatedontheundersideofthe mattressbaseundereachbedend Ona profilingbed securethestrapsaroundthe moveablesectionsofthebase Ensur...

Page 10: ...dhanging hooks Pullthehooksbytherubbertabsto preventaccidentallytrappingyourfingers Ensureitissecurebeforeuse failuretodoso couldresultinequipmentdamage 6 Connecttheumbilicalconnectorstothe socketsons...

Page 11: ...flash green Allow upto 20minutes for completeinflation Donotlieaperson orany weight onthemattressduringinitialinflation Donotplaceaperson oranyitem onthe mattressduringinitialinflation 9 Wheninitialin...

Page 12: ...to feelthe cellsforinflation 2 Withthepatientlyingsupine unzipone sideofthe top cover justpastsacralregion lowerspine 3 Slideyourhandunderneaththepatientand feelfor a deflatedcellunderthepatient s lo...

Page 13: ...therapyanduseradjustablepressuresettings TheselectedmodeisindicatedbyanLEDindicator D AlarmIndicator ThethreeamberLEDslightupin5different patternstoindicatedifferentfaultsthesystem maybeexperiencing...

Page 14: ...Thisfeaturehelpstopreventtheriskofbottoming out Agreen AUTOFOWLER lightilluminateswhenthis functionisactive Itisautomaticallydeactivated whenthemattressbackrestisprofiledbelow5 HPower StandbyandServic...

Page 15: ...ruptionto therapy due to power failure Note Itwilltake16 20hourstofullychargethe battery K BedExitAlarm PressingthebuttonturnsBedExitAlarmonandoff TheONlightabovebuttonilluminatestoindicatethe functio...

Page 16: ...icallydisengage Backup Battery Thelithium ionbackupbatteryisdesignedto providetemporaryuninterruptedtherapywhen mainspowerisunavailableegduringpatient transferorinapoweroutage Thecontrolunitautomatica...

Page 17: ...ksareheard Airwillremaininthe systemforupto24hours dependingonpatient andenvironmentalcircumstances Specialised Heel Therapy ThespecialisedHeelCairzoneenablestheseven narrowalternatingaircellsinthehee...

Page 18: ...sedfromthesystem detachthemattressfromthebedby unfasteningthestraps thenfoldandroll themattressfromheadendtofootendfor storage 6 Returnallitemstothecustomcarrybagfor safekeeping CPRRELEASE POWER STAND...

Page 19: ...hadryclothtoremoveexcess moisture Asoftbristlednylonbrushcanbeusedto gentlycleancrevices Ensurethecontrolunitisdisconnectedfrom mainspowerbeforecleaning Donotspraydisinfectantdirectlyontothe controlun...

Page 20: ...lunitfacedownonasoft flat surfacewithbackpaneluppermost useasoft clothtopreventscratches 3 Useasmallscrewdrivertocarefullyremovethe airfiltercover Cleanthedustfromthefilteror discardandreplacewithanew...

Page 21: ...for kinks obstructions or damage Undo any kinks and obstructions Air cells have failed to reach operating pressure after turning on Turn of control unit and disconnect from power Reconnect to power a...

Page 22: ...ate no display lights Check control unit is connected to the mains power supply Check for loose power cord connection and ensure main power is switched on Check the fuses in control unit Replace if ne...

Page 23: ...ENT 0 4 0 2A FUSE RATING T2AL250V MIN MAX PRESSURE 20 60mmHg 6mmHg FLOW RATE Max10L min PROTECTION TYPE ClassIITypeBF INGRESS PROTECTION RATING IP21 CONTROL UNIT DIMENSION HEIGHT 326mm WIDTH 275mm DEP...

Page 24: ...dsThefirst digit indicates thelevelofprotectionthattheenclosure providesagainst accessto hazardous parts e g electricalconductors moving parts andtheingressofsolidforeignobjects SecondDigit LiquidsPro...

Page 25: ...nsequipment Fortransmittersratedatamaximumoutputpowernotlistedabove therecommendedseparationdistanced inmetres m canbeestimatedusingtheequationapplicabletothefrequencyofthetransmitter wherePisthe maxi...

Page 26: ...ative humidity should be at least 30 Air 2 4 8 15 kV Air 2 4 8 15 kV Electrical fast transient burst IEC 61000 4 4 2kV for power supply lines 2kV for power supply lines Mains power quality should be t...

Page 27: ...eryInUse LEDisnormal duetopowerissuppliedbyinternalbatteryduringpoweroutage Thebackupbatterycouldbeusedforupto3hours EMI EMC Statement and Manufacturer s Declaration Immunitytest IEC60601testlevel Com...

Page 28: ...mendedseparationdistance calculatedfromtheequationapplicable tothefrequencyofthetransmitter Radiated RF IEC 61000 4 3 10V m 80MHz 2 7GHz 80 AMat1kHz 10V m 80MHz 2 7GHz 80 AMat1kHz Recommendedseparatio...

Page 29: ...on Pleasedonotuseanyothercablesoraccessoriesnotapprovedbythemanufacturerinthismanualtoavoidnegative influenceonelectromagneticcompatibility EMI EMC Statement and Manufacturer s Declaration Testfrequen...

Page 30: ...ished good of the same type rather than being repaired Refurbished parts may be used to repair goods NovisHealthcarewillnotbeheldresponsibleforanyrepair otherthanthosecarriedoutbyitoroneofitsAuthorise...

Page 31: ...31 ProCair Trio User Guide Notes...

Page 32: ...Pressure care and patient handling specialists 06102020F 1300738885 novis novis com au 12 12MarsRd LaneCoveWest NSW2066...

Reviews: