ProCair Trio User Guide
System Preparation
7
Feed power cord through the cord retention
fl
aps along either side of the mattress
base. Ensure all press studs are fastened.
Insert power cord plug into the side of the
control unit, then connect to an appropriate
electrical outlet and switch on mains power.
The
Standby
indicator on the control unit will
glow amber, con
fi
rming the control unit is
connected to a power source.
Ensure the power cord is not under strain; is
free from obstruction; and is secured safely so
as not to be a trip hazard.
8
On the control unit, press and hold the
Power
button for a minimum of three
seconds. The
Power
indicator will
glow green to indicate the system is
operational and automatically in
fl
ating.
While in
fl
ating, the
Pressure
Setting
indicators will
fl
ash blue
and the
Max In
fl
ate
indicator
will
fl
ash green. Allow up to 20 minutes for
complete in
fl
ation.Do not lie a person (or any
weight) on the mattress during initial in
fl
ation.
Do not place a person (or any item) on the
mattress during initial in
fl
ation.
9
When initial in
fl
ation is complete, the
Pressure Setting
and
Max In
fl
ate
indicators
will extinguish, to indicate the system is
ready for use. The system automatically
defaults to
Alternating Mode
at start up,
with AutoCair feature (automatic weight
detection) always in operation to select an
appropriate pressure for patient support
according to mattress load, unless
overridden manually.
CPR RELEASE
POWER
STANDBY
ALTERNATE
STATIC
PRESSURE
SETTING
ProCair Trio
LOCATED AT
MATTRESS
HEAD END -
TWIST TO OPEN
NORMAL
<30%
FLASHING: CHARGING
FLASHING: LOW BATT
FLASHING
SERVICE DUE
LOW PRESSURE
SOFTER
NORMAL
FIRMER
HIGH PRESSURE
STARTUP FAILURE
ALTERNATION
FAILURE
SENSOR
DISCONNECTED
ALARM GUIDE
LOCKED
ON
AUTO
FOWLER
BED EXIT ALARM
see alarm guide on left
ALARM INDICATORS
BATT.
IN USE
POWER
3 SEC
3 SEC