www.novasina.ch
44
45
ENGLISH
ENGLISH
„Maintenance“->“Configuration (Export, Import)“ -> Export to SD-card“ and
confirm prompting by [Yes]. If a configuration of the current instrument is already
stored on the SD-card, file is overwritten.
12.2.2 Load Configuration
To import (or load) a configuration,
•
Go to „System settings“-> „Maintenance“->“Configuration (Export, Import)“
-> Import from SD-card“ .
•
Now choose the appropriate configuration. If any configuartion file is detec-
ted on the SD-card, an error message is displayed.
•
Confirm question by [Yes] that configuration is loaded and existing configura-
tion will be overwritten on the meter in following pop-up window.
•
Confirm that system will be booted and new configuartion will be loaded
Note:
•
If a configuration can not be loaded due to corrupt file or while created by a
non-compatible software, system will load the last working version once boot
process is performed. This reset will be notified by an issued error message
after booting.
•
It is possible that configuration created with a newer firmware are not wor-
king on instrument with previous firmware versions.
•
The license code will not be overwritten once configuration is loaded.
12.3 Reset to Factory Settings
It can be necessary that a reset to factory settings need to be performed. This is
done by selecting “System settings“ -> „Maintenance“ ->“Reset to factory set-
tings“ -> „[Yes] -> „[No].
Please note the following:
•
Calibration data and other data on the sensor remain
• Audit Trail remains as well and reset is recorded
• User management and Sample Database will be deleted
•
12.4 System Update
Novasina is offering system updates from time to time (not every month, maybe
once a year or so). Those updates are free of charge.
Please contact your local Novasina distributor and ask about the latest LabMas
-
ter-aw neo update file. Unce received, please proceed as follows:
•
Unzip the received file named update.zip.
•
Copy unziped update file into main directory of the SD-card by using a PC
•
Insert SD-Card into LabMaster-aw neo.
•
Start update process by go to „System settings“ -> „Maintenance“ -> „System
Update“ and press [OK].
•
The update is performed automatically
- First, master software is written into flash memory and system will be
restarted.
- Second, sub-systems (chamber control, humidity and IR sensor) will be
updated.
Note: