12. Additional Functions
The LabMaster-aw neo is equipped with many additional functions to ease
instrument‘s use and handling. Some of the functions are not equally useful for
each user.
12.1 Acoustic and Stable Signals
Various options are available to customize the stable signal appearance. it is also
possible to connect an external optical or acoustic indication (flashing light,..)
12.1.1 Acoustic Signals
To change or modify the accoustic signal settings, go to „System settings“ ->
„Acoustic and stable signals“. The following settings can be made:
•
Volume key beep: Volume of beep if hitting a button
• Volume stability signal: Volume of beep once measurement has reached
stability
•
Duration Stability signal beep: Duration of beep signal once measurement
has reached stability. If no value is entered, beep is switched off.
•
Reminder stability signal beep: Time period, after which stable beep will be
repeated. The repetition is done until beeper is switched off (confirmed) by
a user or measurement chamber has been opened. An active reminder is
displayed in the status bar as well. If „Duration stability signal beep“ is set to
„Off“, this menu point is not displayed.
12.1.2 External Signal
If the LabMaster-aw neo is used in noisy environment, stable notification might
not be recognized. Thereby, an optical or acoustic, external transducer can be
connected to the instrument and be configured to provide a recognisable signal
in such environment. The connector is a 230VAC socket
Once transducer is connected, go to „System settings“ -> „Acoustic and stable
signals“ -> „External stable output“. The following settings can be done:
•
Off: External signal is switched off
• During stable: External singal is switched on as long as stable signal is dis
-
played
•
During stability beep: External signal is switched on fro the time the stable
signal beep is active on the instrument
12.2 Save and Load of Configuration (SD-Card)
Instrument specific data like profiles, sample database and other settings except
calibration data can be exported to the SD-card. To do so, select „System set-
tings“-> „Maintenance“->“Configuration (Export, Import)“.
The exported configuration is stored in the main directory of the SD-card as a
XXXXXXX.cfg file whereas XXXXXXX represents the serial number of the instru-
ment. By this labeling, it is possible to store configurations of different instru-
ments on one SD-card.
Of course it is possible by this to transfer a once set configuration of one Lab-
Master-aw neo to other LabMaster-aw neo to create several instruments with
exact same configuartion.
To protect the exportet files and to avoid that passwords can be read-out, files
are AES-256 Bit encrypted.
12.2.1 Save Configuration
To export (or save) a configuration, simply proceed to „System settings“->