B-2
Prime CCS Instructions for Use Manual
C
x
E
=
E
x
-
E
std =
S log
____
Equation 4
C
std
By holding C
std
in Equation 4 constant, E is dependent on
only 1 variable, C
x
, the concentration of the ion of interest
in the sample. Equation 5 can be rearranged to isolate this
variable:
C
x
= (C
std
) 10
(E/S)
Equation 5
The analyzer’s microcomputer uses Equation 5 to calculate
the concentration of sodium, potassium, calcium, and
chloride ions in the sample.
B.1.2 pH Sensor
Definition of pH
The pH of an unknown sample is calculated using the
following equation:
E
std C
– E
x
pH E
std C
- E
x
= pH
std C
+ ––––––––– Equation 6
Slope
E
std C
– E
x
where Slope =
——————————
Equation 7
pH
std C
– pH
std D
Principle of pH Measurement
pH is measured using a hydrogen ion selective membrane.
One side of the membrane is in contact with a solution of
constant pH. The other side is in contact with a solution
of unknown pH. A change in potential develops which is
proportional to the pH difference of these solutions.
This change in potential is measured against a reference
electrode of constant potential. The magnitude of the
potential difference is a measure, then of the pH of the
unknown solution.
Summary of Contents for Stat Profile PRIME
Page 1: ...Stat Profile PRIME CCS Analyzer Instructions for Use Manual 52920...
Page 2: ......
Page 4: ......
Page 6: ......
Page 20: ...1 10 Prime CCS Instructions for Use Manual...
Page 67: ...A 7 Appendix A Appendix A...
Page 68: ...A 8 Prime CCS Instructions for Use Manual...
Page 69: ...A 9 Appendix A Appendix A...
Page 70: ...A 10 Prime CCS Instructions for Use Manual...
Page 71: ...A 11 Appendix A Appendix A...