2 Getting Started
2.3 Automatic Calibrations
The Stat Profile Prime CCS analyzer performs a 2-point
calibration on all analytes and air detectors 30 minutes
after being powered on and regularly thereafter to maintain
optimal sensor card and air detector performance. A 1-point
calibration is performed at regular intervals to monitor the
sensor card’s performance between each 2-point calibration.
If a calibration error occurs, an alert is shown to notify
the operator and the test button of the affected analyte is
displayed with an
orange
background to indicate it is not
available for testing.
Scheduled 2-point calibrations can be delayed once for 10
minutes by pressing the Cancel button. After 10 minutes the
rescheduled calibration will begin and cannot be cancelled.