3-1
3. Anal
ysis
3 Sample Analysis
3 Sample Analysis
When Ready is displayed on the Home screen the analyzer
is ready to analyze samples for any analyte not displaying
an
Orange
test button. The analyzer can measure whole
blood samples from capillary tubes, syringes, test tubes, and
open containers as well as external Quality Control material
from ampules and internal Quality Control material from an
internal QC cartridge
3.1 Analyzing Patient Samples
Before running a patient sample verify the analyzer is Ready
to perform an analysis and that all the desired analytes are
available for selection. If necessary, refer to Chapter 2 for
additional information.
3.1.1 Analyzing Syringe Samples
From the Home screen, login if you are prompted to login.
1. Press the
Login
icon to log into the analyzer.
2. Enter or scan your
Operator ID
then press
.
3. If required, enter or scan your
Password
then press
.
Figure 3.1 Operator ID Screen and Operator Password Screen
If login is not required or after logging into the analyzer:
1. Select the syringe
icon from the container
drop-down list.