6-4
Prime CCS Instructions for Use Manual
6.3.3 Flushing the Sample Probe
1. From the Home screen, press the Tool Box
icon.
From the System Menu select Replace Probe and wait
for pump to stop.
2. Disconnect the air detector's sample line from the
Reference Sensor module using the removal tool
attached to the door.
3. Use the syringe with attached tubing to flush distilled
water through the S-line and out the Sample Probe.
A small sample cup can be held in front of the
Sample Probe to catch the water.
4. With the Sample Probe now cleared, reconnect the
S-line to the Reference Sensor
5. Close the door.
6. Press the Calibrate
icon.