6-2
Prime CCS Instructions for Use Manual
6.2 Error Log
From the Service Menu, select Error Log. The Error Log
screen is displayed with errors displayed in chronological
order. To go to the next or previous error log pages, press
Next Page or Previous Page. To print the error log, select
Printer Menu from the Service Menu.
Table 6.1 lists the analyzer’s error logs and the corrective
action.
Table 6.1 Stat Profile Prime Error Logs
Error
Corrective Action
To Be Determined
6.3 Operator Flow Test
The flow test verifies that fluid can be pulled through the
system from the probe. If water cannot be pulled through
the system, a clog or leak exists. The procedure for the flow
test and sensor cartridge, reference sensor, and sample
probe back flush are as follows.
6.3.1 Flushing the Reference Sensor
1. From the Home screen, press the Tool Box
icon.
2. From the System Menu select the Change Reference
Sensor and wait for the pump to stop.
3. Open the door; open Cartridge door.
4 Remove Sensor Card.
NOTE: To change the Reference Sensor the Sensor Card
must be removed first. Only touch Sensor Card by
the edges.
5. Unplug pump tubing manifold from top of sensor.
6. Disconnect the S-line from the bottom of the sensor
using the removal tool attached to the door.
Summary of Contents for Stat Profile PRIME
Page 1: ...Stat Profile PRIME CCS Analyzer Instructions for Use Manual 52920...
Page 2: ......
Page 4: ......
Page 6: ......
Page 20: ...1 10 Prime CCS Instructions for Use Manual...
Page 67: ...A 7 Appendix A Appendix A...
Page 68: ...A 8 Prime CCS Instructions for Use Manual...
Page 69: ...A 9 Appendix A Appendix A...
Page 70: ...A 10 Prime CCS Instructions for Use Manual...
Page 71: ...A 11 Appendix A Appendix A...