5 Periodic Replacements
5.2 Reference Sensor Replacement
1. From the Home screen, press the Tool Box
icon.
2. From the System Menu select the Change Reference
Sensor and wait for the pump to stop.
3. Open the door; open Cartridge door.
4 Remove Sensor Card.
NOTE: To change the Reference Sensor the Sensor Card
must be removed first. Only touch Sensor Card by
the edges.
Figure 5.8 Reference Sensor
Pump Tubing
Manifold
S-Line
Sensor Card
Reference
Sensor
Reference
Sensor Arm