This D/P transmitter has a specified accuracy of ±0.15%. To 
ensure your calibration maintains this degree of accuracy, 
the equipment used to check zero and span settings must 
conform to the recommendation listed below. Use of 
equipment with less than specified accuracy figures will not 
provide acceptable results. King assumes no responsibility 
for transmitter accuracy if test equipment does not meet 
recommended minimum standards.

Recommended Equipment:

MILLIAMMETER;

 3-1/2 digit minimum, 0.05% accuracy 

(20 ohms max. internal resistance)

DC POWER SUPPLY

PRESSURE INDICATOR; PSI or IN. OF WATER

0.05% accuracy corresponding to transmitter range
5   psid nominal 

± .0025 psi or .0693" of water

10 psid nominal 

± .005  psi or .1385" of water

15 psid nominal 

± .0075 psi or .2078" of water

30 psid nominal 

± .015  psi or .4156" of water

50 psid nominal 

± .025  psi or .6927" of water

NOTE

—The low pressure port (“E”) of transmitter must be 

vented to atmosphere during the calibration and adjustment 
procedures. 

0

160

Compressed

Air

Precision

Pressure

Regulator

Precision

Pressure

Indicator

S

20.00

10 A

300 mA

COM

OFF

VOLTS AC

VOLTS DC

300 mV DC

OHMS

AUDIO

AMPS DC

AMPS AC

Meter

4-20 mA

Output

Connection

+

+V

A

V

COM

R

L

+

+

Summary of Contents for KING-GAGE 5900

Page 1: ...e to Range Calculation Calibration Procedures Zero Span Adjustments for Differential Pressure Transmitters Model 5900 D P Module 4 mA to 20 mA Pressure Transmitter Multiple Tank Level Display KING GAGE A NOSHOK Company I 1010 West Bagley Road Berea Ohio 44017 I Ph 855 367 2494 I www king gage com ...

Page 2: ...ties of fitnessforaparticularpurposeormerchantability aregiven bythisagreement Theexclusiveremedyfornonconformity of these goods shall be repair and or replacement of the nonconforming goods or parts Sellerwillnotbeliableforconsequentialdamagesresulting from breach of this agreement The term consequential damages shall include but shall not be limited to damage toallmachines equipmentandgoodsother...

Page 3: ...pan Adjustment 6 Low Pressure Input for differential pressure 8 Calibration Test Setup Typical 9 Pressure Range Calculation 10 Troubleshooting 11 5900 1 2 Fittings 1 none 1 8 NPT connection 4 Swagelok 1 4 tube 1 8 NPT compatible with copper or poly tubing Range 0 0 5 psid 0 0 3 bar 1 0 10 psid 0 0 69 bar 2 0 15 psid 0 1 03 bar 3 0 30 psid 0 2 06 bar 4 0 50 psid 0 3 44 bar Model No Designations ...

Page 4: ...r Supply Voltage 14 40 Vdc unregulated Power Supply Stability Effect on FSO less than 0 005 of span change in output per volt change at input terminals Burst Pressure 200 psi Span Adjustment Range Transmitter adjustment turndown is 3 1 for the 0 15 0 30 0 50 psid nominal ranges Nominal Minimum Range of Adjustment 0 5 psid 0 3 psid 0 83 in thru 0 138 in water 0 2 1 m thru 0 3 5 m water 0 10 psid 0 ...

Page 5: ...Pressure Transm itter T E S T Plug in Power Signal Connector Ð 2 1 RL K IN G G A G E ZERO SPAN D P M O DULE 4 2 0 m A Pressure Transm itter T E S T Plug in Power Signal Connector Ð 2 1 Power Requirement The D P transmitter requires a source of DC power 14 40 Vdc for operation Minimum excitation voltage must be no less than 14 Vdc Any receiver installed on the signal loop meters data loggers contro...

Page 6: ...t to the maximum depth hydrostatic pressure Measure the output using the meter terminal posts on the transmitter board 2 Adjust span potentiometer as necessary to increase output clockwise ordecreaseoutput counterclock wise until the signal displays as 20 00 or 20 01 mA on the test meter 3 Next step is to re check the zero output and slightly adjust if necessary In Process Zero Span Adjustment It ...

Page 7: ...lease refer to the individual DIP switch positions listed for adjusting the existing setting of the transmitter The four 4 switches provide 16 increments of coarse adjustment Prior to changing any switch setting check the current output of the transmitter at the desired upper pressure value Also note the existing switch setting 1 Apply the desired maximum pressure to the sensing diaphragmoftheD Pt...

Page 8: ...tter sensing element and the high pressure to the other side the internal tank pressure above the liquid low pressure is balanced across the sensing element This is also referred to as being equalized which is why we designate the low pressure connection as E CAUTION While the internal pressure above the liquid is not a factor in determining range span there may be instances where the sum of the l...

Page 9: ...ipment MILLIAMMETER 3 1 2 digit minimum 0 05 accuracy 20 ohms max internal resistance DC POWER SUPPLY PRESSURE INDICATOR PSI or IN OF WATER 0 05 accuracy corresponding to transmitter range 5 psid nominal 0025 psi or 0693 of water 10 psid nominal 005 psi or 1385 of water 15 psid nominal 0075 psi or 2078 of water 30 psid nominal 015 psi or 4156 of water 50 psid nominal 025 psi or 6927 of water NOTE ...

Page 10: ...rve represents the distance from the lowest point on the bottom of the tank to the installed sensor Full is the level of contents at which the tank is filled to desired level which is less than the actual full capacity of the tank The liquid contents in the tank has a specific gravity of 1 032 C 1 032 Specific Gravity B 300 Upper Level A 28 Reserve 300 28 1 032 27 6807 10 14 psid 300 28 1 032 281 ...

Page 11: ...20 0 mA Check zero output of transmitter adjust zero to 4 00 mA if necessary Check span setting adjust to 20 00 mA at nominal range or calculated full pressure Maximum output less than 20 00 mA Check that applied pressure is within span limits for transmitter range Transmitter has 40 span adjustment Check span setting adjust to 20 00 mA at nominal range or calculated full pressure Higherresistivel...

Page 12: ......

Reviews: