Flexineb MDI Use (Fig. 2):
1. Remove the Controller, Cable and Medication Cup
2. Plug the supplied Adaptor into the Inlet Valve on top of the Aerosol Chamber
3. Rotate the Aerosol Chamber to the horizontal position
4. Insert the MDI and administer medication following the advice of your veterinarian
Controller Functions
Operation LED:
Green
= System On
Constant Red=
Low Battery Warning, typically
time left for one treatment before re-charging is
required.
Flashing Red
= Battery is dead and you must re-
charge the battery to continue using the system.
Warning LED:
LED Flashes and Controller Powers Off
means
there is a break in the electrical circuit. The
problem may be a disconnected or broken cable
or broken connection in the medication cup.
Check the connections or replace the broken
cable or medication cup to fix the problem.
LED Constant On
means an internal electronic
fault has occurred and the unit needs to be
returned for repair.
Contact your Customer Service line for advice.
>-
'
/
�lexineb
®
E
2
Battery Charging LED:
Green
means the Flexineb
1-+-
•
.
.....
,...
charger is connected and
6
-
•
0
the battery is charging.
lJ'
!
r
The Green LED will turn off
when the battery has
reached full charge.
Full charging can take 4 to 5
On
I
Off Switch
hours if the battery is
completely discharged.
If the battery cannot reach
full charge the LED wi II not
turn off and the battery may
need replacing.
A battery that cannot reach
full charge will give reduced
operation time.
Technical Information
Flow Rate:
Typically :?:1.0 ml/min with 0.9% NaCl solution. The flow rate is very dependent on the drug used.
Aerosol Droplet Size:
Typically 70% of particles generated are less than 5 microns VMD measured using Malvern Spraytech with
0.9% NaCl solution. The aerosol droplet size is dependent on the drug used.
Medication Cup Volume:
The medication cup will hold 10ml of liquid but is expandable to 30ml using the plug in extension cup supplied.
The controller has a timer function therefore, if the extension cup is used the controller will need to be
re-started to complete the dose.
Battery Power:
Battery Type: 7.4V Li-ion rechargeable battery. Use the Flexineb supplied battery only. Use of another type of
battery may present a risk of fire or explosion.
Operating Time: >3 hours from fully charged.
Battery Charging:
Use the Flexineb supplied charger only. Use of an alternative charger may present a risk of fire or explosion. A
fully discharged battery can take 4 to 5 hours to reach full charge.
Mains Power Supply:
(I
nput:
100
to
240
v AC
/
50
to
60
H
z
)
The Flexineb can also be operated using the Power Supply provided with the system. The same Power
Supply charges the battery. Use the Flexineb Power Supply only.
Use of an alternative Power Supply may present a risk of fire or explosion and void warranty.
Environmental Conditions:
Operating: 5
°
C to 45
°
C and up to 95% RH
Storage and Transportation: -20
°
C to +45
°
C and
15% to 95% RH
•