42
42 Keyboard tare available on channel 1 only.
43 Tare greater than capacity
44 Invalid Password number range, can only be 0000-9999.
45 Parameter 1 to enable password, 0 to disable
46 Only a value of 0, 1 or 80 is accepted as a parameter
47 *
48 Invalid Filter value
49 Invalid Filter Fast step value
50 Invalid Fast step Sensitivity
51 Invalid Fast step on/off
52 Invalid Tare Function Parameter 0-4
53 Invalid input for AD voltage range
90 Calibration checksum failed
100 SRAM failure
110 RTC RAM failure
112 Clock Reset
115 Clock Failed
120 Battery flat or does not exists
121 Battery must be removed
130 COM1 loop back test failed
131 COM2 loop back test failed
133 COM driver chip failed
140 FLASH memory erase failed
141 FLASH memory write failed
151 Database CRC failed
152 CAL copy CRC failed
153 Ticket Buffer CRC failed
150 Audit trail CRC failed
154 DPAGE stack overflow
185 SMART wire COM link not responding SmartWire is enabled and is
trying to find devices. This error will occur as a result of not having devices
connected to the Smart Wire port or a bad wire connection. Hit clear to bypass
this error message. To disable smart wire use parameter 59 in weight mode.
Enter 0 to disable smart wire and 1 to enable it again.
186 SMART wire set-point checksum failed
191 Channel 1 AD converter not responding
192 Channel 2 AD converter not responding
193 Channel 3 AD converter not responding
Error message “Cannot Print”
Note: If you try and print and there is motion, or the scale is overweight then
a
message will scroll across the display “Cannot Print”. It is important to
understand