The range of each programming section can be modified. To do so, change the respective
>Switchover height< accordingly.
Specify positioning values accurately
Observe to accurately specify the positioning values of the individual sensors.
Sensors with built-in pressure measurement cell shall be installed at the lowest point of the
channel bottom (sensor Type CSP-V2D, CSP-V2U and CSM-V1D).
If sensors should feature an elevated mounting position (block or similar) or should be in-
stalled on a berm, it is necessary to enter a different value as mounting height. Measure the
distance from the
bottom edge
of the sensor mounting plate down to the lowest point of the
application.
Then enter this value as >Mounting height<.
The
centre
of the sensor mounting plate is the reference value in the event of tilted or in-
clined sensors (such as lateral installation within pipes or trapezoid channels).
The sensor position value is accepted automatically for the position of the v-sensor as well as
for the determination of v-crit (see chapter “36.3.5 v-Determination low levels”).
It is possible to additionally enter an offset to adjust the pressure sensors.
Fig. 36-12 Programming integrated pressure sensor
The same specifications apply for flow velocity sensors with integrated water-ultrasonic sen-
sor. Measure the distance from the top edge of the sensor crystal.
The system automatically detects and considers these differences accordingly while connect-
ing the sensors.
36.3 Setting Parameters in Menu v-Sensors
Apart from the measurement place and the level sensors, the third relevant point is setting
the parameters of the flow velocity sensors. Except type and number of sensors this menu
moreover includes the spatial position. The values specified within this menu relate to the de-
fined channel regarding shape and spatial dimension (see chapter “36.1.2 Channel Profiles”).
36.3.1
Number of Flow velocity sensors
You can connect one flow velocity sensor to the NivuFlow Mobile 750 transmitter.
The option to connect up to three flow velocity sensors is currently not supported, however, in
preparation.
Summary of Contents for NivuFlow 750
Page 141: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 141 Approvals and Certificates...
Page 142: ...Instruction Manual NivuFlow Mobile 750 page 142 NFM 750 rev 03 27 05 2020...
Page 143: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 143...
Page 144: ...Instruction Manual NivuFlow Mobile 750 page 144 NFM 750 rev 03 27 05 2020...
Page 145: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 145...
Page 146: ...Instruction Manual NivuFlow Mobile 750 page 146 NFM 750 rev 03 27 05 2020...
Page 147: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 147...
Page 148: ...Instruction Manual NivuFlow Mobile 750 page 148 NFM 750 rev 03 27 05 2020...
Page 149: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 149...
Page 150: ...Instruction Manual NivuFlow Mobile 750 page 150 NFM 750 rev 03 27 05 2020...
Page 151: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 151...
Page 152: ...Instruction Manual NivuFlow Mobile 750 page 152 NFM 750 rev 03 27 05 2020...
Page 153: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 153...
Page 154: ...Instruction Manual NivuFlow Mobile 750 page 154 NFM 750 rev 03 27 05 2020...
Page 155: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 155...
Page 156: ...Instruction Manual NivuFlow Mobile 750 page 156 NFM 750 rev 03 27 05 2020...
Page 157: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 157...
Page 158: ...Instruction Manual NivuFlow Mobile 750 page 158 NFM 750 rev 03 27 05 2020...
Page 159: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 159...
Page 160: ...Instruction Manual NivuFlow Mobile 750 page 160 NFM 750 rev 03 27 05 2020...
Page 161: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 161...
Page 162: ...Instruction Manual NivuFlow Mobile 750 page 162 NFM 750 rev 03 27 05 2020...