30 / 32
scd340en21p07
5.4.9.
Error codes
Error Code
Error description
Causes and actions to be done
Err 1
Memory error
Contact local agent
noEc
or
Err 2
Echo loss or echo too weak to evaluate
Bad reflection from the surface, reflection not directed towards the sensor,
due to dusting excessive sound absorption. Check device selection and/or
adjustment
Err 3
Hardware error
Contact local agent
Err 4
Calculation overflow
Check settings
Err 5
Code referring to sensor error or improper installation/mounting, level
in the dead band
Verify sensor for correct operation and check for correct mounting according
to User’s Manual
Err 6
The measurement is at the reliability threshold
Change aiming or try to find a better location
Err 7
No signal received within the measuring range specified in P04 and
P05.
Review programming, also look for installation mistake
Err 12
Linearisation table error: L(1) and L(2) are both zero (no valid data-
pairs)
See the Section ”Linearisation”
Err 13
Linearisation table error: there are two same L(i) data in the table
See the Section “Linearisation”
Err 14
Linearisation table error: the r(i) values are not monotone increasing
See the Section “Linearisation”
Err 15
Linearisation table error: measured Level is higher than the last
Volume or Flow data-pair
See the Section “Linearisation”
Err 16
The checksum of the program in the EEPROM is wrong
Contact local agent
6.
MAINTENANCE, REPAIR
The device does not require regular maintenance. In case dust adheres to the face of the sensor despite the self-cleaning of the sensor face through resonance,
(e.g.: static build-up) it can be cleaned by pressurized air.
The warranty card contains the terms and conditions. Before returning the device for repairs, it must be cleaned thoroughly. The parts in contact with the medium
may contain harmful substances; therefore, they must be decontaminated. Our official form (
Returned Equipment Handling Form
) must be filled and enclosed in
the parcel. Download it from our website
. The device must be sent back with a declaration of decontamination. A statement must be provided in
the declaration that the decontamination process was successfully completed and that the device is clean from any hazardous substances.
7.
STORAGE CONDITIONS
Ambient temperature: –30…+60 °C (
−
22…+140 °F)
Relative humidity: max. 98 %