P65: - - - - Maximum temperature of the transducer [°C]
P66: - - - - Minimum temperature of the transducer [°C]
P70:
Number of Echoes / Echo Map (separate menu item or highlighted window)
Entering this parameter the level transmitter stores the actual echo map, the amplitude and the position of the echoes.
The distance and amplitude of these echoes can be read-out one by one in ascending order. The selected echo is marked.
Note:
for graphic display of the echoes use the EView software or a MultiCONT process controller.
P71: - - - - Distance of the of Measuring Window (DIST) [ m ]
P72 - - - -
Amplitude of the selected echo [dB] ( < 0 )
If the echo map is empty,
NoECHO
is displayed.
P73: - - - -
Position of the selected echo (time)
ms
P74: - - - -
Signal to Noise Ratio
R
ATIO
M
EASUREMENT CONDITIONS
Over 70
Excellent
Between 70 and 50
Good
Under 50
Unreliable
P75: - - - - Blocking Distance
This parameter gives information about the actual close-end blocking distance (provided automatic blocking was selected in
P05)
5.4.8.
Test parameters
P96: - - - -
Checksum
P97: - - - -
Software code
P98: - - - -
Hardware code
P99: - - - - Access lock by secret code
The purpose of this feature is to provide protection against accidental programming of parameters or by a person. not entitled. The secret code
can be any value other than
0000
. Setting a secret code will automatically be activated when the EasyTREK is returned to the Measurement
Mode. In order to program locked device, the secret code should be entered first in
P99
. Thus, for entering a new code or erasing the old one the
knowledge of the previous code is necessary.