P06: - - - -
Far-end blocking
FACTORY DEFAULT: 0
Far-end blocking is the range below the level set in parameter
P06
. Far-end blocking can be used to avoid disturbing effects
of stirrers or heaters at the bottom of the tanks.
When the unit detects echoes in this range it gives special
signals.
When level sinks below far-end blocking:
-
Message of "
Sub 0
" appears in EView
(in level and volume measurement mode)
- Current output holds value corresponding to far-end blocking
distance
When level rises above far-end blocking:
The calculation of level and volume will be based on the programmed tank dimensions, therefore the measured or calculated process values
will not be influenced in any way, by the value of far-end blocking.