Instruction Manual: EMP-3
NIC-TD-0000067-01
Nippon Instruments Corporation
71
6.5.4.4 CALIBRATION PROCEDURE
(1)
Press “MENU” and then “Maintenance”.
(2)
Proceed to the “Calibration/Manufacturer” screen.
(3)
Enter the password.
(4)
The following screen will be displayed. Press “Calibrate”.
(5)
Press the numerical button.
(6)
Enter a standard gas concentration to be used for calibration, and press “Enter”.
Button
Description
Cancel
Cancellation of alteration
Clear
Deletion of all characters
Enter
Determination
BS
Deletion of one character