Chapter 1
Assembly
20
Ass
embly
17
Attach the eyelevel riser (optional).
Tool: Hex driver (2 mm across flats)
Place the eyelevel riser on the arm, and tighten the fixing
screws on the front of the arm using a hex driver (2 mm
across flats).
Securing the eyelevel riser
18
Attach a camera and connect the DS camera control unit (optional).
18.1
When attaching to the DSC port for the ergonomic binocular tube
Tool: Hex driver (2 mm across flats)
(1) Screw the camera head into the C mount on the
DSC port.
(2) Remove the rear cover of the ergonomic tube and
insert the DSC port.
(3) Secure the DSC port in place with the DSC port
fixing screw using the tool provided with the
microscope.
(4) Connect the camera connection connector for the
camera head and the camera connector for the
DS-L3/DS-U3 DS camera control unit using a
provided camera cable.
(5) Connect the EXT.I/O connector for DS-L3/DS-U3
with the DSC connector for the microscope or the
Ex-OUT connector for the connector box B using a
camera trigger cable.
For DS-L3, a USB cable can also be used for
connecting the USB (H) connector for DS-L3 and
the USB connector for the connector box B.
Precautions for connecting the camera
trigger cable
When connecting the camera trigger cable with the
DSC connector, be sure to insert it completely.
Prior to photomicrography, adjust the camera
position as appropriate. (See Chapter 3 “18
Capturing Images
-
Photomicroscopy” in the
“Operation”)
Attaching a camera to the DSC port
for the ergonomic binocular tube
Eyelevel riser fixing
screw
Eyelevel riser
DSC port
fixing screw
Camera connection
connector
Camera head
DSC port