Chapter 1
Assembly
5
Ass
embly
2
Attach the DIA motorized shutter (optional).
The motorized shutter is attached by Nikon.
Contact your nearest Nikon representative when the DIA motorized shutter needs to be attached or removed.
3
Attach the lamp.
Tool: Hex wrench (3 mm across flats)
Lamp handling precautions
Avoid touching the glass surface of the lamp with
your bare hands.
(1) Loosen the lamphouse cover fixing screw and lift up
the cover to remove.
(2) Insert the lamp clamp lever to open the socket pin
hole. Attach the lamp while holding down the lever.
Put the lamp clamp lever back to its original position.
Designated lamp: PHILIPS7724
or OSRAM HLX64623
(3) Reattach the cover back to its original position and
tighten the lamphouse cover fixing screw.
Removing the lamphouse cover
Attaching the lamp
4
Attach the lamphouse for dia-illumination.
Tool: Hex driver (2 mm across flats)
(1) Insert the lamphouse at the rear of the microscope
and tighten the fixing screws.
(2) Connect the dia-illumination lamphouse cable to the
LAMP connector on the rear of the main body.
DSC
LAMP CTRL
(BOX B
)
LAMP
DC12V 100W
MODEL ECLIPSE Ni-U
NIKON CORPORATION
TOKYO, JAPAN
100–240V~
1.7A
50/60Hz
M A D
E I N
J A P
A N
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that may cause undesired operation.
This Class A digital apparatus complies with Canadian ICES-003.
Cet appareil numérique de la classe A est confirme à la norme NMB-003 du Canada.
9 2 0 0
0 1
INSPECTION
EQUIPMENT
4N75
1.
2.
3.
Do not touch the lamphouse while the lamp is lit.
The surface of the lamphouse becomes hot when the
lamp is on.
Turn off the power and allow the lamp and lamohouse
to cool enough before replacing the lamp.
Wait for at least 30 minutes after turning off the
lamp.
Use 12V100W HALOGEN lamp only.
HALOGEN 12V100W
9 5 2 0
0 1
NI-LH
CAUTION!
-High Temperature-
JAPAN
Securing the dia-illumination lamphouse
Lamphouse
fixing screw
Fixing screw
Lamphouse
cover
Insert the hex wrench and
loosen the fixing screw.
Lamp
clamp lever
Lamp
Lamphouse fixing screw