Chapter 3 Part Names and Microscopy Procedures(With DIH-M Mounted)
3.4 Epi-fluorescence Microscopy
48
13
To return to bright-field
microscopy.
•
Close the shutter of the Epi-illumination and
block the Epi-illumination.
•
Rotate the excitation method switchover
turret to move the position where no filter
cube is installed into the optical path.
•
If the condenser is not mounted, install the
condenser, then focus and center the
condenser. (Refer to step 12 of bright-field
microscopy.)
•
Turn on the microscope power switch to turn
on the diascopic light source.
14
Turn off all power switches after
completing observations.
Press the power switch
to the "
◯
" position.
Use the excitation method switchover
turret to change cubes.
Press the switch to the
"|" position.
Close the shutter.