Additional page
JAA79051-R.3624.A
- L28-6・ AF-S DX18-70/3.5-4.5G -
January. 28. 2005
⑧
Unlock the holder-moving lever, and move the holder down slowly by the lever. Insert 3 alignment screw-
drivers (long, middle, short) in the loosened screw holes of the 5th lens chamber.
Note)
Because the screws cannot be seen, when inserting the alignment screwdrivers, put them straight down in
the screw holes so that the screws can be easily found.
⑨
Turn the power of the Monitor, LINE GENERATOR, and MEGALIGHT 100 to ON. By checking the
screen of the Monitor, rotate the micrometers (X and Y) so that the intersection point of the cross lines*
1
(calibrated) comes in the center of the Monitor.
*
1
Adjust the cross lines by turning the "LIGHT CONT." knob of "MEGALIGHT 100" and the focus
ring until the calibration of the cross lines can be seen clearly.
⑩
When the intersection point of the cross lines comes in the center, tighten the 3 screws of the 5th lens
chamber with the alignment screwdrivers.
Note) When the holder is raised, the intersection point of the cross lines is misaligned. So make an
adjustment by considering this.
⑪
Move the holder-moving lever up slowly to lock the holder, and remove the lens from the equipment (for periphery).
LIGHT CONT. knob
Focus ring
Cross line
Note) In case the cross lines are tilted, adjust them
by turning the chart, which is screwed in the rear
tube of the equipment. (Just slight slackness of the
chart poses no problems.)
INC