54
Return to previous page
5.7.6 Measurement does not start even when start button is pressed
Activate manual mode.
Is there conduction between the P902 1st and 2nd
pins on the joystick (30601-EA92) when the start
button is pressed
?
No
Replace the Joystick ASSY (30601-2600) (see
7.1.1 [p69]).
Yes
Are there any breaks in the U/D switch cable
(30601-EA25)
?
Yes
Replace the U/D switch cable (30601-EA25).
No
Are there any breaks in the BA02-BA03 cable
(30601-EA24)
?
Yes
Replace the BA02-BA03 cable (30601-EA24).
No
Replace the Driver board (30601-BA03) (see 7.2.3
[p75]).
Is the problem resolved?
No
Replace the Main board (see 7.2.1 [p73]).
Yes
End