5.2.32 ERR123 (Chart initialization error)
5.2.33 ERR501 (Power error of Eye Care card)
5.2.34 ERR502 (Writing error of Eye Care card)
Are there any breaks in the chart (30703-EA74)
Yes
Replace the Chart (30703-EA74) (see 7.3.15
[p92]).
No
Replace the Driver board (30601-BA03) (see 7.2.3
[p75]).
Is the problem resolved?
Yes
End
No
Replace the Main board (see 7.2.1 [p73]).
Replace the Eye Care card.
Is the problem resolved?
No
Replace the Base board (30601-BA05) (see 7.2.4
[p76]).
Yes
End
Reinsert the Eye Care card for writing.
Does ERR502 appear?
Yes
Replace the card.
No
End