96
Return to previous page
8 . Attach the tension jig (30670-2100) to the
measuring unit ASSY (30601-3000).
9 . Turn FC4 × 10 of the tension jig (30670-
2100) to adjust the tension of the wire
(30601-M307) so that the focus base
ASSY (30601-3300) moves smoothly.
10. Fasten the pulley 2 ASSY (30601-3120)
with BS3 × 6 (n = 2).
7.3.20 Switching power supply (80602-00102)
Replacement part: 80602-00102
1 . Remove the bottom plate (30601-M002)
2 . Unscrew BS3 × 8 (n = 4) to replace the
switching power supply (80602-00102).
3 . Reassemble the parts in the reverse order.
$5
(%
$5
Summary of Contents for AR-310A
Page 2: ......
Page 10: ...XAR17ARDA001G E 10 Return to previous page This page is intentionally left blank ...
Page 62: ...XAR17ARDA001G E 62 Return to previous page This page is intentionally left blank ...
Page 68: ...XAR17ARDA001G E 68 Return to previous page This page is intentionally left blank ...
Page 102: ...XAR17ARDA001G E 102 Return to previous page This page is intentionally left blank ...
Page 186: ...XAR17ARDA001G E 186 Return to previous page 9 3 1 2 30601 E000 AR 310A and ARK 510A ...
Page 190: ...XAR17ARDA001G E 190 Return to previous page 30601 EA14 ...
Page 192: ...XAR17ARDA001G E 192 Return to previous page 30601 EA23 30601 EA24 ...
Page 194: ...XAR17ARDA001G E 194 Return to previous page 30601 EA30 30601 EA31 30601 EA32 ...
Page 196: ...XAR17ARDA001G E 196 Return to previous page 30601 EA51 30601 EA52 ...
Page 217: ......