52
Return to previous page
5.7.5.2 Alignment is not performed in up-and-down direction
Does an error appear when the power is turned
on?
Yes
5.2 Error Message Appears (p22)
No
Is the auto tracking icon displayed?
No
Display the auto tracking icon.
Yes
Is the mire ring displayed? (Check it by using a
model eye.)
No
5.7.2 Mire ring is not displayed (p49)
Yes
Are there any breaks in the BA02-BA03 cable
(30601-EA24)
?
Yes
Replace the BA02-BA03 cable (30601-EA24).
No
Are there any breaks in the BA01-BA03 cable
(30601-EA13)
?
Yes
Replace cable BA01-BA03 (30601-EA13).
No
Does the voltage between the A5 and B5 pins of
P204 (J4) on the tracking board (30601-BA02)
change in the range of DC ±5 V (polarity is
reversed according to the directions) when the
alignment is moved in the up-and-down direction
from the focused position?
No
Replace the Tracking board (30601-BA02) (see
7.2.2 [p74]).
Yes
Summary of Contents for AR-310A
Page 2: ......
Page 10: ...XAR17ARDA001G E 10 Return to previous page This page is intentionally left blank ...
Page 62: ...XAR17ARDA001G E 62 Return to previous page This page is intentionally left blank ...
Page 68: ...XAR17ARDA001G E 68 Return to previous page This page is intentionally left blank ...
Page 102: ...XAR17ARDA001G E 102 Return to previous page This page is intentionally left blank ...
Page 186: ...XAR17ARDA001G E 186 Return to previous page 9 3 1 2 30601 E000 AR 310A and ARK 510A ...
Page 190: ...XAR17ARDA001G E 190 Return to previous page 30601 EA14 ...
Page 192: ...XAR17ARDA001G E 192 Return to previous page 30601 EA23 30601 EA24 ...
Page 194: ...XAR17ARDA001G E 194 Return to previous page 30601 EA30 30601 EA31 30601 EA32 ...
Page 196: ...XAR17ARDA001G E 196 Return to previous page 30601 EA51 30601 EA52 ...
Page 217: ......