When
is clicked, displayed tables and peak profiles are cleared so that a new
measurement can be made.
Item
Description
Clear both STD and SAMPLE
data
All table items for standard and unknown samples are cleared.
Clear SAMPLE data
All table items for an unknown sample are cleared.
Use this for a measurement reusing a calibration curve.
Clear all measurement data
Table items only on the right of the absorbances of standard and
unknown samples are cleared.
Use this when the calibration curve creation and unknown sample
measurement are carried out in fixed procedures.
Clear SAMPLE measurement
data
Of the measurement values of unknown samples, table items only
on the right of absorbances are cleared.
Use this when a calibration curve is reused and the unknown
sample measurement is carried out in a fixed procedure.
* When a check mark is placed for “Clear MEMO”, the memo entered in the table condition
is deleted.
* When a check mark is placed for “Clear Self Check result”, the instrument check result is
cleared.
Click
and open and display a file in which a measurement result has been
saved.
Select a file to be opened and click “Open”.
Click
and save measurement data.
Enter a file name and click “Save”.