Instruction Manual: Thermal Vaporization Mercury Analyzer MA-3
Solo
NIC-TD-0000090-15
Nippon Instruments Corporation
115
Warning
When touching the instrument for maintenance, make sure that it
has been cooled sufficiently.
When touching, put on protectors such as heat-resistant gloves.
Otherwise, you may get burned.
Disconnect the power cable from the instrument before
maintenance work.
Otherwise, an electric shock or a failure caused by a short circuit
may result.
Caution
The inlet is locked while the instrument is turned off.
Do not open the inlet while it is locked.
The inlet can be damaged.
When touching a glass component, take care not to suffer an
injury such as cutting a hand.
(1)
Loosen the nut of the reducing joint connected to the mercury collector tube, and
remove the reducing joint.
(2)
Loosen the screw of the sample heating tube retainer, and remove it.
(3)
Turn the inlet knob counterclockwise while pushing up manually the solenoid lock
locking the boat in/out rod.
(4)
Pull out the boat in/out rod.
(5)
Remove the screw fixing the boat in/out rod, and remove it.
Step (1)
Step (2)
Step (3), (4)
Turn the inlet knob and pull out while
pushing up the solenoid lock manually.
Step(5)