Flow in Measurement Path
Using the on/off control of the valves V1 through V3, three patterns of a flow in the
measurement path are used during a measurement.
The V1 is turned on, the V2 and V3 are turned off and ordinary air flows through the path
indicated with the red lines.
Water produced in the thermolysis of sample is removed by the drain bottle, and gas is
discharged through the purge line.
Heater H3
(Mercury Collector Tube
Heating Furnace)
Mercury
Collector Tube
Heater H2
(Catalyst Furnace)
Heater H1
(Sample Heating Furnace)
Heater H4
Detector
Absorption
Cell
Lamp
Detector
Drain Bottle
Humidifica-
tion Bottle
Needle Valve
Air Pump Flow Sensor
Activated
Carbon Filter
Activated Carbon Filter
Nafion Tube
Exhaust
Air
Filter Unit