Instruction Manual: Thermal Vaporization Mercury Analyzer MA-3
Solo
NIC-TD-0000090-15
126
Nippon Instruments Corporation
In the case of “AD FULL”, “LAMP BROKEN” or “NG” for the “REF” or “SIG” voltage in
the instrument check, make the voltage adjustment of the optical system.
For the details of the screen, refer to “5.6.3 Optical Adjust”.
(1)
Click the “Maintenance” tab in the upper part of the screen, and then click
.
(2)
Click “AUTO” in the lower right part of the displayed screen, and the adjustment of
the optical system will be automatically made.
Warning
When touching the instrument for maintenance, make sure that it
has been cooled sufficiently.
When touching, put on protectors such as heat-resistant gloves.
Otherwise, you may get burned.
Disconnect the power cable from the instrument before
maintenance work.
Otherwise, an electric shock or a failure caused by a short circuit
may result.
When tubes are replaced because of low sensitivity or an abnormal blank value or when
a component is replaced and tubes around it are also replaced, replace necessary tubes
(colored ones) referring to the following figures:
Piping Diagram (Top View) 1
1
2
3
6
7
8
4
5
9
10
11