Instruction Manual: Thermal Vaporization Mercury Analyzer MA-3
Solo
NIC-TD-0000090-15
132
Nippon Instruments Corporation
Warning
When touching the instrument for maintenance, make sure that it
has been cooled sufficiently.
When touching, put on protectors such as heat-resistant gloves.
Otherwise, you may get burned.
Disconnect the power cable from the instrument before
maintenance work.
Otherwise, an electric shock or a failure caused by a short circuit
may result.
(1)
Remove the connector of the air pump.
(2)
Remove the three screws of the air pump retainer.
(3)
Loosen the four screws fixing the air pump, and remove and replace the air pump.
(4)
Attach the components in the procedure reverse to that for the removal.
(5)
When the air pump has been replaced with a new one, clear “Air Pump Usage” of
the GLP counter.
For details, refer to “5.6.2 GLP Counter”.
Step (1)
Step (2)