Hunan Next Generation Instrumental T&C Tech. Co., Ltd.
NGI
l
NGI
N36100 Series
Wide Range High Voltage Programmable DC Power Supply
27
4
V_MO
Voltage monitoring output
5
GND
Digital ground
6
EX_ISET
Current programming input
7
GND
Digital ground
8
EX_VSET
Voltage programming input
After APG-V function is activated, the output voltage is controlled by externally input
voltage signal. The external programming reference voltage is 0~5V. 0~5V
programming signal corresponds to 0 ~ full scale output voltage.
Let’s take model N36190-150-12 for example. The input analog value is 0 ~ 5V, and
the output voltage is between 0 ~ 150V. When the input analog voltage increases by
1V, N36100 output voltage increases by 30V. When the input analog voltage is 3V,
N36100 output voltage will be 90V.
7.7 Protection
Steps to enter
Protection
:
Press
Menu
button→ Select
Protection
by rotating
→ Press
Enter
button.
On Protection interface, users can set parameters of over voltage, over current and
over power. When protection occurs, alarm information will be shown on the screen.
Please press
Shift
button first and then
ESC
button to clear the protection manually.
Figure 26 Protection Setting Interface
Note: Only voltage programming input is valid of N36100 analog programming.
Note: To exit from
APG
, please press
Menu
button or the function button required.