Newton Scientific
EVO 10 Centrifuge
|
14
Location
The centrifuge is only to be operated indoors. The set-up location must fulfill the following
requirements:
•
A safety zone of at least 30 cm must be maintained around the centrifuge. People and
hazardous substances must be kept out of this safety zone while centrifuging.
•
The supporting structure must be stable and free of resonance.
•
The supporting structure must be suitable for horizontal setup of the centrifuge.
•
The centrifuge is not to be exposed to heat and strong sunlight.
WARNING
UV rays reduce the stability of plastics.
Do not subject the centrifuge, rotors and plastic accessories to direct sunlight.
•
The set-up location must be always well-ventilated.
Transporting/Moving
WARNING
Always ensure all rubber foot are off the ground by unscrewing using a spanner. Once the wheels
are in contact with the floor, you can move the centrifuge by rolling. Never lift the centrifuge by its
front panel, its back panel or at its door. Always remove the rotor before moving the centrifuge.
Due to its weight, the centrifuge should be controlled/moved by at least two people.
To prevent possible injuries at least two people should lift and carry the centrifuge by holding it at
the bottom from opposite sides.
Transport the centrifuge and accessories upright within the associated packaging, if possible.
NOTICE
Store the original centrifuge packaging. Contact a shipping company or the technical support for the
transport.
Always remove the rotor before moving the centrifuge. If you do not remove the rotor you might
damage the centrifuge drive or centrifuge spindle.