EVO 10 Centrifuge
|
15
Setting up & Leveling
Once the centrifuge is in its final position screw the rubber feet entirely down. First use your hand
then the wrench to lower the feet. The casters should be 5-15 mm off the ground.
Level the centrifuge as follows:
1.
Place the enclosed bubble level on the top surface of the rotor.
2.
Adjust the centrifuge feet until the bubble inside the level is completely inside the marked
area.
3.
Turn the rotor with the bubble level around a full 360°.
When 50 % of the bubble remains within the marked area, the leveling of the centrifuge is OK. If
more than 50 % of the bubble leaves the marking, the centrifuge must be leveled again
CAUTION
If the centr
ifuge isn’t leveled, imbalances can occur and the centrifuge can be damaged.
Never place
anything under the centrifuge feet to level the centrifuge.
The centrifuge is to be placed on horizontal and level supporting structures or floor. Horizontal level
is to be checked when moving the centrifuge to a new location.
Mains Connection
NOTICE
Only connect the centrifuge power cable to grounded electrical socket/isolator.
1.
Turn off the power supply switch when not in use. Power supply switch is on the right-hand
side of the EVO 10 centrifuge.
2.
Check whether the cable complies with the safety standards of your country.
3.
Make sure that the voltage and frequency correspond to the figures on the rating plate.
4.
Establish the connection to the power supply with the supplied power cable.