Newton Scientific
EVO 10 Centrifuge
|
28
Troubleshooting Guide
If problems occur other than those listed in this table, the authorized technical support
representative must be contacted.
General Errors
Causes
Actions
Display does not light up when
the power is switched on
No power supply to the
centrifuge
Check cable connections.
Check there is power at
socket.
Blown fuse
Replace fuse.
Abnormal vibration during run Rotor is off balance
Check the loading of the rotor
and load the rotor
symmetrically.
Tube rupture
Visual inspection of tubes.
Rotor is not tightened
Tighten the rotor nut by
turning clockwise using the
supplied rotor key.
Swing-rotor buckets does not
swing smoothly
Clean the pivots and grooves
and apply grease.
Damage shock absorber
Contact a service technician.
Centrifuge run cannot be
started
Control system error
Switch the centrifuge off and
on again. If the problem
persists, contact a service
technician.