Newton Scientific
EVO 10 Centrifuge
|
20
Acceleration / Deceleration Rates
The Newton Scientific Evo 10 offers you 10 selections for each plus the ability to have the option to
set to “0” brake
. The setting is displayed above the ACCEL / DECEL button on the touchscreen.
Press the ACCL / DECL button to key in the desired rates. Press Enter to confirm. The last profile is
saved when you restart the centrifuge.
Pre-Selecting Speed/RCF
RPM stands for Revolutions Per Minute.
RCF stands for Relative Centrifugal Force and allows better transfer of protocols between centrifuges
and rotors of differing size.
Ensure that the rpm or RCF is correctly set.
1.
Press the RPM/RCF button to select RPM or RCF.
2.
Enter the desired value by using the number keypad and press Enter to confirm. RPM / RCF can
be viewed during a run on the top left-hand corner of the screen.
NOTICE
The minimum motor speed is 150 rpm. Any extremely low RPM settings will be automatically
increased to the minimum RPM at 150 rpm.
Running Time Pre-Selection
1.
Press the Time display window to access the number keypad.
2.
Enter the desired runtime in hh:mm:ss.
3.
Press Enter to confirm.
Preselecting the Temperature
You can preselect temperatures between -20 °C and +40 °C. To set the temperature, proceed as
follows:
1.
Press the TEMP display area to access the number keypad.
2.
Key in the desired temperature.
3.
Press Enter to confirm.