86
Remote Command Set
TE?
Description
Error code query
Syntax
TE?
Remarks
This command is used to read the error code. The error code is one numerical
value up to three(3) digits long. (see Appendix for complete listing) In general,
non-axis specific errors numbers range from 0-99. Axis-
1
specific errors range
from
1
00-
1
99, Axis-
2
errors range from
2
00-
2
99 and so on.
Note:
Errors are maintained in a FIFO buffer ten(10) elements deep. When an
error is read using TB or TE, the controller returns the last error that occurred and
the error buffer is cleared by one(1) element. This means that an error can be read
only once, with either command.
Response
Value
Description
nn
integer
Error code
Example
TE?
(Get error status)
108
(Returns 108; motor 1 not connected)
TE?
(Get error status)
0
(Returns 0; no error detected)
See Also
TB