92
Remote Command Set
HOSTNAME
Description
Hostname set.
Syntax
HOSTNAME
name
Remarks
This command is used to set the controller’s hostname. A hostname is a label or
nickname assigned to a device connected to a computer network and is used to identify
the device in various forms of communication. This feature enables users to assign the
controller a more easily identifiable name, perhaps linked to its actual function in an
application (e.g., TIPTILT, MIRROR-2). This command is more common with Ethernet
but can also be used with USB communication.
NOTE: Send this command via USB interface as Ethernet communication may
become unstable during Ethernet settings configuration. While connected via USB,
use the Picomotor application software to send commands (Go to ‘Terminal’ tab).
The default hostname for the controller is 8742-
serialnum
, where
serialnum
is the
controller’s serial number. The hostname can have a maximum of 15 alphanumeric
characters and hyphen (“-”) character. The first character in the hostname cannot be a
hyphen (“-”). Issue Save (SM) and Reset (RS) commands to controller for this parameter
to take affect.
Argument
Value
Description
name
string
Hostname
Example:
HOSTNAME TIPTILT-1;SM;RS
(Set hostname, save, and reset system)
HOSTNAME?
(Get hostname)
TIPTILT-1
(Returns “TIPTILT-1”; hostname)
See Also
IPMODE, SM, RS
HOSTNAME?
Description
Hostname query.
Syntax
HOSTNAME?
Remarks
This command is used to query the controller’s hostname
Response
Value
Description
name
string
Hostname
Example:
HOSTNAME?
(Get hostname)
TIPTILT-1
(Returns “TIPTILT”; hostname)
See Also
IPMODE
Summary of Contents for New Focus Picomotor 8742
Page 1: ...Picomotor Controller Driver User s Manual Model 8742 ...
Page 53: ...Computer Interfacing 53 Cycle Tab Cycle Motion ...
Page 54: ...54 Computer Interfacing Terminal Tab Command Line Terminal ...
Page 57: ...Computer Interfacing 57 Options Window Connection Options Setup ...
Page 58: ...58 Computer Interfacing Properties Window USB and Ethernet Properties ...
Page 59: ...Computer Interfacing 59 Controller Scan Scan RS 485 LAN ...
Page 61: ...Computer Interfacing 61 Device Address Window View Change Controller Address ...