32
ENTERING OR CHANGING YOUR (DEFAULT) ID CODE:
Enter
ID
menu (Repeat steps from 1 to 3).
Press
d
10)
Display advances to
CH.ID
Change ID Code Submenu.
Press
d
11)
Display shows
0000
message with flashing first digit.
If you want to change your default “ID Code” you can do it now,
otherwise press
a
and menu will skip to
FULL
Full Security
Submenu.
Press
b
&
c
12)
Press
b
and
c
buttons to enter your new “ID Code”
number.
Press
d
13)
Display shows
STRD
stored message momentarily and
then advances to the
FULL
Full Security Submenu.
FULL SECURITY LEVEL SUBMENU:
Press
d
14)
Display flashes
ENBL
Enable or
DSBL
Disable.
Press
b
15)
Scroll through the available selections: “Enable” or
“Disable”.
Press
d
16)
Display shows
STRD
stored message momentarily and
then advances to
SP.ID
Setpoint/ID Submenu.
If “Full” Security Level is “Enabled” and the user attempts to enter
the Main Menu, they will be prompted for an ID Code. The ID Code
should be correct to enter the instrument Menu item.
SETPOINT/ID SECURITY LEVEL SUBMENU:
This Security Level can be functional only if
FULL
Security Level is
Disabled.
Press
d
17)
Display flashes
ENBL
Enable or
DSBL
Disable.
Press
b
18)
Scroll through the available selections: “Enable” or
“Disable”.
Press
d
19)
Display shows
STRD
stored message momentarily and
then advances to
COMM
Communication Submenu.
If
“Setpoint/ID”
Security Level is
“Enabled”
and the user attempts
to advance into the
CNFG
Configuration Menu, he will be prompted
for ID Code number. The ID Code should be correct to proceed into
the Configuration Menu, otherwise display will show an Error and
skip to the Run Mode.
If
“Full”
and
“Setpoint/ID”
Security Levels are “
Disabled
”, the ID
code will be “Disabled” and user will not be asked for ID Code to
enter the Menu items (“ID” Submenu will not show up in “ID/Setpoint”
Menu).
Summary of Contents for i16-AL
Page 17: ...13 3 2 Menu Configuration Figure 3 1 Flow Chart for ID and Set Points Menu...
Page 45: ...NOTES...
Page 46: ...NOTES...
Page 48: ...M5290 N 0613...