![Neuropace NeuroPace CL-315-10 User Manual Download Page 56](http://html1.mh-extra.com/html/neuropace/neuropace-cl-315-10/neuropace-cl-315-10_user-manual_1676338056.webp)
Programming Instructions
54
RNS
®
System User Manual
P
ROGRAMMING
I
NSTRUCTIONS
L
OGGING ONTO THE
N
EURO
P
ACE
®
P
ROGRAMMER
1. Connect the wand to the USB port on the programmer.
2. Connect the AC adapter to the programmer and to the power cord.
3. Plug the power cord into a grounded power outlet.
4. Turn on the programmer by pressing the power button.
5. Enter your
U
SER
N
AME
and
P
ASSWORD
.
6. Click
OK
to login.
T
ESTING THE
W
AND
S
IGNAL
1. Go to the programmer
M
AIN
M
ENU
screen.
2. Place the wand over the neurostimulator within approximately 3 cm of the device.
3. Select the
T
EST
W
AND
S
IGNAL
button.
The
S
IGNAL
L
EVEL
(S) measures the strength of the communication between the neurostimulator and
the programmer.
The
S
IGNAL
Q
UALITY
(Q) measures the absence of electronic interference in the environment.
WARNING: D
RAINING THE
N
EUROSTIMULATOR
B
ATTERY
Testing the wand placement over the neurostimulator for more than 10 minutes
per day may drain the neurostimulator battery prematurely.
I
NTERROGATING THE
RNS
®
N
EUROSTIMULATOR
Interrogating the RNS
®
Neurostimulator is the act of retrieving the currently programmed settings and
any stored data from the neurostimulator using the programmer. The wand must be held in place over
the neurostimulator for interrogation to succeed.
Note: After each session during which the neurostimulator was interrogated by the programmer, it
is recommended that the programmer be synchronized with the PDMS.
Interrogating the Neurostimulator:
1. Go to the programmer
M
AIN
M
ENU
screen.
2. Select the
I
NTERROGATE
button.
Two vertical bars indicate signal level (S) and signal quality (Q) as described in