Responsive Neurostimulator System
23
RNS
®
System User Manual
WARNING: C
OMPATIBILITY WITH SIMILAR IMPLANTABLE PRODUCTS
The NeuroPace
®
RNS
®
Neurostimulator, NeuroPace
®
Cortical Strip Lead, and
NeuroPace
®
Depth Lead are not compatible with non-NeuroPace leads and / or pulse
generators. Incompatible configurations may cause damage to the products and may
result in unsafe current densities delivered to the patient.
L
EAD
C
OMPONENT AND
A
CCESSORIES
Tunneling Tool (Model TT-01), Tunneling Tool Tip (Model TTT-01), Tunneling Straw (Model TTS-01)
The tunneling tool components are used to tunnel an implanted lead from its cranial exit point,
through a sub-galeal pathway, to the implanted RNS
®
Neurostimulator location.
Lead Cap (Model LC-01)
The lead cap protects the proximal end of a lead when it is not connected to the RNS
®
Neurostimulator.
Stop Gauge (Model SG-01)
The stop gauge is placed on a depth lead prior to implantation to indicate the appropriate depth of its
insertion.
Suture Sleeve (Model SS-01)
The suture sleeves protect the lead body when sutures are used to secure a lead.
N
EURO
P
ACE
®
P
ROGRAMMER
NeuroPace
®
Programmer (Model PGM-300)
The programmer includes a laptop personal computer that runs software developed by NeuroPace,
and uses custom telemetry components to communicate with the RNS
®
Neurostimulator by holding
the wand over it. The wand is a applied to the patient (an applied part).
The programmer is used to prescribe how the neurostimulator operates. Settings include, but are not
limited to, the implanted system configuration, detection settings adapted to the patient's ECoG
patterns
,
ECoG record storage, and the settings of the therapies.
The NeuroPace
®
Programmer may also be used on its own to review previously retrieved
neurostimulator activity information, perform detection analysis, and to communicate with the
NeuroPace
®
Patient Data Management System (PDMS) via connection over the Internet.