Surgical Procedures
35
RNS
®
System User Manual
R
ECOMMENDED
I
MPLANT
P
ROCEDURE
F
LOW
C
HART
Make scalp incision at the
desired implant location
Trace the Craniectomy Template
on the skull at the implant location
Neurostimulator Implant
Make craniectomy around
template drawing
Secure the Ferrule
in the craniectomy
Perform appropriate craniotomy
to implant Cortical Strip Lead
Separate and resect dura
Implant Cortical Strip
Lead subdurally
Stabilize Cortical Strip Lead using
Suture Sleeves as appropriate
Attach Burr Hole Cover
base to skull
Advance insertion cannula
to appropriate depth
Advance Depth Lead
to target depth
Secure Depth Lead using
Burr Hole Cover Cap
Make Burr Hole at appropriate
location to implant Depth Lead
Cortical Strip Lead Implant
Depth Lead Implant
Tunnel Leads to Neurostimulator
Place the excess Lead under the scalp in a pocket area near the burr hole or
craniectomy area. Do not place the Leads on or around the Neurostimulator
Insert the Lead(s) into the Connector Cover and secure the cover
onto the Neurostimulator. A Connector Plug must be placed in the
Connector Cover if only one Lead is implanted.
Secure the Neurostimulator in the Ferrule
Using the Programmer, confirm the integrity of the system
Record the Lead serial numbers and Neurostimulator
port numbers in the Programmer
Close incisions over the burr hole, craniotomy, and Neurostimulator