Surgical Procedures
45
RNS
®
System User Manual
4. Advance the lead from the burr hole to the craniectomy, allowing the proximal lead contacts
and lead body to extend from the craniectomy.
5. Remove the tunneling tool straw.
Note: Leave sufficient slack in the lead adjacent to skull fixation points to relieve stress on the lead.
Securing the Neurostimulator and Lead(s)
1. Hold the connector cover with the windows visible.
2. Place the lead proximal ends into each port of the connector cover until the lead stops.
3. Confirm that the lead marker band is flush with the edge of the connector cover. If there is a
gap between the band and the edge of the connector cover seal, gently push the lead into the
connector cover port again until the lead stops.
4. Confirm that the lead contacts are fully visible in each window of the connector cover (Figure
14).
Figure 14: Connector cover with proximal lead contacts in place.