Surgical Procedures
49
RNS
®
System User Manual
4. Clear tissue and bone debris away from the area surrounding the neurostimulator and lead
strain relief area.
5. Remove the lead strain relief from the neurostimulator and dispose.
6. Use the torque driver to loosen and remove the connector cover.
7. Remove the proximal lead ends from the connector cover. Make sure the proximal lead ends
are appropriately marked so the leads can be connected to correct ports of the new
neurostimulator. Dispose of the old connector cover.
Precaution: Lead Damage
Bending, kinking, and stretching of the lead may cause lead damage. Handle
the lead with care.
8. Using the torque driver, partially loosen the screw in the ferrule clamp and swivel the clamp
90 degrees away from the neurostimulator to facilitate removal of the neurostimulator.
9. Remove the neurostimulator from the ferrule.
Note: Return the explanted neurostimulator and leads to NeuroPace. NeuroPace will
provide shipping containers if requested.
10. Flush the craniectomy and ferrule.
11. Open the new neurostimulator package.
Note: The neurostimulator is packaged with a temporary white cover over the lead contacts.
Leave this cover in place until you are ready to attach the leads to protect the
contacts during handling.
12. Place the new neurostimulator in the ferrule.
13. Secure the neurostimulator to the ferrule by rotating the ferrule clamp into position over the
neurostimulator and tightening the ferrule clamp screw until the torque driver clicks (Figure
12).
Note: DO NOT use any tools other than the torque driver provided to secure the ferrule
clamp in place.
14. Remove the white temporary cover from the neurostimulator.
15. Flush the neurostimulator connector area (if necessary) to remove debris and dry area before
attaching the new connector cover.
16. Hold the new connector cover with the windows visible.
17. Place the lead proximal end into each port of the new connector cover until the leads stop.
18. Confirm that the lead marker band is flush with the edge of the connector cover. If there is a
gap between the band and the edge of the connector cover seal, gently push the lead into the
connector cover port again until the lead stops.
19. Confirm that the lead contacts are fully visible in each window of the connector cover (Figure
14).