User Manual DigiMano 2500
XXIV
Maintenance and Storage
Recommended Procedures and Precautions
Maintenance of DigiMano 2500 is straightforward, requiring little more than keeping it
clean. It is important to keep dust and dirt from getting inside the unit or into the inlet
ports.
Following precautions are to be observed while using the unit.
Do not exceed the maximum rated pressure. Although a transient pressure
(hammer pressure) of three times the rated maximum can generally be tolerated
without recalibration, prolonged exposure to higher pressure will cause sensor
damage and recalibration may be required.
Do not connect directly to any strong corrosive gases or fluids.
Do not connect pressure line to the vacuum port ‘
Port 2
’ (dual port Models)
Do not use when the low battery warning comes on.
Make sure the built in LI-Ion polymer is fully charged prior to use.
Do not use any other charger other than the one provided with the unit.
Calibration / Service
The mechanical assembly of DigiMano 2500 contains no parts that can be serviced by
the user. The unit should be returned to Netech Corporation for repair or calibration. The
alignment and adjustment parameters are critical to the robust and efficient performance
of the unit and can be performed only at the factory.
The unit is factory calibrated with NIST traceable standards and recommended to be
calibrated once a year.
Netech maintains a complete repair and recalibration service at a very low cost and fast
turnaround. Estimates for repair and recalibration are available upon request.
DigiMano 2500 contains NO USER SERVICEABLE PARTS and calibration/ service
should be performed only by Netech. Attempt to repair / service the unit outside
Netech voids the warranty.