Tap
Display
D
MENU (Displays 1 second).
DEV TYPv.
ENT
RTV CNTv.
D
K FACT A (Displays 1 second).
######## (Current value blinks).
ENT
K FACT B (Displays 1 second).
######## (Current value blinks).
ENT
ROv SPV.
Press B, Reset to 0, and Total now increase along with the register.
ENT
ATv MANv.
Press D for Manual Reset.
ENT
DEC LOC.
ENT
SmartTrol returns to normal operating mode.
Table 33 – Incrementing Total
Start the flow or reattach the register to the meter. As the register advances, so does Total
as displayed on the SmartTrol.
You may need to perform this procedure periodically due to rounding errors in the K
Factor. To improve the matching of the percentage difference between Total and the
register, adjust the K Factor by the percentage difference between Total and the register
reading.
The adjustment is determined by the following formula:
Example:
l
10-inch Trident Turbine K Factor = .16
l
Start register reading = 106400
l
SmartTrol Total reading = 106400
After some period of operation:
l
Current register reading = 408000
l
Current SmartTrol Total = 408257
TRICON
®
SmartTrol
™
Installation and Maintenance Guide
67